Página 1 - Data Acquisition Unit

User’sManualMX100Data Acquisition UnitIM MX100-01E5th Edition

Página 2 - Revisions

1-1IM MX100-01EExplanation of Functions32145Index1.1 Overview of the MX100The MX100 consists of the main module equipped with an Ethernet port, input

Página 3 - Safety Precautions

4-12IM MX100-01E4.5 10-CH, Medium-Speed Universal Input Module (MX110-UNV-M10) SpecificationsStyle number: S1Type of measurement: DC voltage, thermo

Página 4 - Software Handling Precautions

4-13IM MX100-01ESpecification32145IndexThe inputs indicated below can be used on MXLOGGER or the MX100/DARWIN API (software sold separately).InputMeas

Página 5 - WARNING

4-14IM MX100-01EMaximum input voltage: DC voltage at 1-V range or less, thermocouple, RTD, and DI (contact): ±10 VDC (continuous) Other measurement

Página 6 - Contents

4-15IM MX100-01ESpecification32145IndexEffects of Operating ConditionsThe specifications below apply when the integration time is 16.67 ms or more.War

Página 7

4-16IM MX100-01E4.6 30-CH, Medium-Speed DCV/TC/DI Input Module (MX110-VTD-L30) SpecificationsStyle number: S3Type of measurement: DC voltage, thermo

Página 8 - Chapter 4 Specication

4-17IM MX100-01ESpecification32145IndexThe inputs indicated below can be used on MXLOGGER or the MX100/DARWIN API (software sold separately).InputMeas

Página 9

4-18IM MX100-01ECommon-mode voltage: 600 VACrms (50/60 Hz), reinforced (double) insulationCommon-mode rejection ratio: For integration time is 16.67

Página 10 - 1.1 Overview of the MX100

4-19IM MX100-01ESpecification32145IndexEffects of Operating ConditionsThe specifications below apply when the integration time is 16.67 ms or more.War

Página 11 - Input/Output Modules

4-20IM MX100-01E4.7 6-CH, Medium-Speed Four-Wire RTD Resistance Input Module (MX110-V4R-M06) SpecificationsStyle number: S2Type of measurement: DC v

Página 12

4-21IM MX100-01ESpecification32145IndexThe inputs indicated below can be used on MXLOGGER or the MX100/DARWIN API (software sold separately).InputMeas

Página 13

1-2IM MX100-01EMain Module (MX100-E)The main module is equipped with a power supply connector, a power switch, an Ethernet port, a CF card slot, and o

Página 14 - Base Plate (MX150)

4-22IM MX100-01ECommon-mode voltage: 600 VACrms (50/60 Hz), reinforced (double) insulationCommon-mode rejection ratio: When integration time is 16.67

Página 15 - PC Software

4-23IM MX100-01ESpecification32145IndexEffects of Operating ConditionsThe specifications below apply when the integration time is 16.67 ms or more.War

Página 16 - 1.2 Main Module Functions

4-24IM MX100-01E4.8 4-CH, Medium-Speed Strain Input Module (MX112) SpecificationsStyle number: S2Number of inputs: 4Input type: Strain gauge or st

Página 17 - Range Over

4-25IM MX100-01ESpecification32145IndexAllowable input voltage: ±10 VDC (between H-L) continuousAllowable common-mode voltage: Channel-to-channel: 3

Página 18 - Remote RJC (RRJC)

4-26IM MX100-01E4.9 10-CH, High-Speed Digital Input Module (MX115) SpecificationsStyle number: S1(-D05), S2(-D24)Input type: -D05: contact (non-vol

Página 19 - Alarm Output Timing

4-27IM MX100-01ESpecification32145Index4.10 8-CH, Medium-Speed Analog Output Module (MX120-VAO-M08) SpecificationsStyle number: S2Number of outputs:

Página 20

4-28IM MX100-01ESetting Value of Span and Preset, and Specified Value of User OutputAnalog output setting value span and preset valueSpecified Value o

Página 21 - Saving Data to the CF Card

4-29IM MX100-01ESpecification32145Index4.11 8-CH, Medium-Speed PWM Output Module (MX120-PWM-M08) SpecificaionsStyle number: S2Number of outputs: 8U

Página 22

4-30IM MX100-01EExternal DimensionsUnit: mm 57131150.8If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the tolerance is

Página 23 - CF Card Specifications

4-31IM MX100-01ESpecification32145IndexSetting by Pulse Interval RangeWhen the pulse interval is 10 msPulse interval PC setting value10 ms 120 ms 2300

Página 24 - Input Module

1-3IM MX100-01EExplanation of Functions32145Index30-CH, Medium-Speed DCV/TC/DI Input Module (MX110-VTD-L30, MX110-VTD-L30/H3) MX110-VTD-L30(clamp t

Página 25 - Common Mode Voltage

4-32IM MX100-01E4.12 Specifications Common to the 8-CH Medium-Speed Analog Output Module and the 8-CH Medium Speed PWM Output Module (MX120)Specifica

Página 26 - Universal Input Module

4-33IM MX100-01ESpecification32145Index4.13 10-CH, Medium-Speed Digital Output Module (MX125) SpecificationsStyle number: S1Number of outputs: 10Co

Página 27

5-1IM MX100-01EDual Save Functions (/DS Option)32145IndexChapter 5 Dual Save Functions (/DS Option)5.1 Overview of FunctionsThe following functions

Página 28 - TC/DI Input Module

5-2IM MX100-01E5.2 Explanation of FunctionsThe PC software programs referred to in this section are the MX100 Standard Software and MXLOGGER (sold se

Página 29

5-3IM MX100-01EDual Save Functions (/DS Option)32145IndexSaving data to the CF CardTransmission of measured data to the PC via communications and savi

Página 30

5-4IM MX100-01EDual Save Function (When Using the PC Software)• Normal Operation BackupFileNo.=0 FileNo.=1InfoBackupONIdleMeasureStartStopBackupO

Página 31

5-5IM MX100-01EDual Save Functions (/DS Option)32145IndexDual Save Function (When Using an Application Based on the MX100/DARWIN API• Application Exa

Página 32 - Measurement Range

5-6IM MX100-01EStarting and Stopping Measurement and RecordingIn situations where the MX100 cannot be connected to a PC, this function can be used to

Página 33

5-7IM MX100-01EDual Save Functions (/DS Option)32145IndexOperation When the CF Card Is Inserted CF Write (Data Write Timing): The timing at which da

Página 34

5-8IM MX100-01E• If a Media Error Occurs When Saving Data to the CF Card ModeDataCF WriteCF FileBackupFileNo.=0InfoMeasureMeasureIdleErrorlostSta

Página 35

1-4IM MX100-01E10-CH, High-Speed Digital Input Module (MX115-D05-H10) • Minimum measurement interval: 10 ms• Maximum number of inputs: 10 inputs• In

Página 36 - Input Module (-D05, -D24)

5-9IM MX100-01EDual Save Functions (/DS Option)32145Index5.3 CF Card Data FileFile NameData files created on the CF card using the 5s-SW or Dual Save

Página 37 - Output Module

Index-1IM MX100-01EIndexDdata viewer software... 1-6DC voltage/current standard...

Página 38

Index-2IM MX100-01EMmain module ... 1-2, 1-7maintenance/test ...

Página 39 - PWM Output Module

Index-3IM MX100-01EUunit number... 1-11upper limit alarm ...

Página 40 - Explanation of Functions

1-5IM MX100-01EExplanation of Functions32145Index10-CH, Medium-Speed Digital Output Module (MX125-MKC-M10) • Output update interval: 100 ms (shortes

Página 41 - Steady Output Operation

1-6IM MX100-01EPC SoftwareOne MX100 can be connected to a PC, and the MX100 Standard Software (that can acquire measured data) is included. The MX100

Página 42

1-7IM MX100-01EExplanation of Functions32145Index1.2 Main Module FunctionsThe main module is the heart of the MX100. Ethernet portPower switchCF ca

Página 43

1-8IM MX100-01EMeasurement Time (When Using the MX100 Standard Software)When a measurement start request or a measurement data output request is made

Página 44

1-9IM MX100-01EExplanation of Functions32145IndexFiltersThe main module is equipped with a first-order lag filter (see section 2.7, “Measures against

Página 45

1-10IM MX100-01EAlarms This function outputs alarms when measured or computed values* meet certain conditions. Select up to four alarms of the followi

Página 46 - 2.1 Handling Precautions

iIM MX100-01EThank you for purchasing the MX100 Data Acquisition Unit. This user’s manual describes the functions, installation and wiring methods, op

Página 47 - 2.2 Installation

1-11IM MX100-01EExplanation of Functions32145IndexContents Displayed on the 7-Segment LEDThe two-digit 7-segment LED displays the unit number, operati

Página 48 - Attachment to a DIN Rail

1-12IM MX100-01ESaving Data to the CF CardData Saving When Communication Is Disconnected and Dual Save FunctionIf communication is disconnected while

Página 49 - 2.3 Attaching the Modules

1-13IM MX100-01EExplanation of Functions32145IndexPrecautions When Using the CF CardOn the MX100, measured data can only be saved to the CF card (back

Página 50

1-14IM MX100-01ELog File StorageIn addition to the measured data that is saved when the communication with the PC is disconnected, a log file of infor

Página 51 - VextVext

1-15IM MX100-01EExplanation of Functions32145Index1.3 Functions of the 4-CH, High-Speed Universal Input ModuleThis module is equipped with four input

Página 52

1-16IM MX100-01EThe following inputs can be used on MXLOGGER or MX100/DARWIN API sold separately. These inputs cannot be used with the MX100 Standard

Página 53 - Screw Terminal Block

1-17IM MX100-01EExplanation of Functions32145Index1.4 Functions of the 10-CH, Medium-Speed Universal Input ModuleThis module is equipped with ten inp

Página 54

1-18IM MX100-01EThe following inputs can be used on MXLOGGER or the MX100/DARWIN API (sold separately). These inputs cannot be used with the MX100 Sta

Página 55

1-19IM MX100-01EExplanation of Functions32145Index1.5 Functions of the 30-CH, Medium-Speed DCV/TC/DI Input ModuleThis module is equipped with 30 inpu

Página 56 - Wiring Procedure

1-20IM MX100-01EThe following inputs can be used on MXLOGGER or the MX100/DARWIN API (sold separately). These inputs cannot be used with the MX100 Sta

Página 57 - One-Gauge Method

iiIM MX100-01ESafety PrecautionsAbout This Manual• Please pass this manual to the end user.• Read this manual thoroughly and have a clear understand

Página 58 - Adjacent Two-Gauge Method

1-21IM MX100-01EExplanation of Functions32145Index1.6 Functions of the 6-CH, Medium-Speed Four-Wire RTD Resistance Input ModuleThis module is equippe

Página 59 - Opposing Two-Gauge Method

1-22IM MX100-01EThe following inputs can be used on MXLOGGER or the MX100/DARWIN API (sold separately). These inputs cannot be used with the MX100 Sta

Página 60 - Four-Gauge Method

1-23IM MX100-01EExplanation of Functions32145Index1.7 Functions of the 4-CH, Medium-Speed Strain Input Module (-B12, -B35, and -NDI)This module is eq

Página 61 - Wiring with the -D24 Option

1-24IM MX100-01EInitial Balancing (Initial Unbalance Value Adjustment)When configuring a bridge circuit with a strain gauge, the bridge circuit will n

Página 62 - Wiring the PWM Output Modules

1-25IM MX100-01EExplanation of Functions32145IndexScaling Settings When Using a Strain Gauge Type Sensor The following is an explanation of scaling se

Página 63 - Main Output Specifications

1-26IM MX100-01EWhen Listed at the Calibration CoefficientAn example is given with a displacement gauge.• Rated input: 20 mm• Calibration coefficien

Página 64

1-27IM MX100-01EExplanation of Functions32145Index1.8 Functions of the 10-CH, High-Speed Digital Input Module (-D05, -D24)The -D05 module is equipped

Página 65

1-28IM MX100-01E1.9 Functions of the 8-CH, Medium-Speed Analog Output ModuleThis module is equipped with 8 outputs that can output voltage or current

Página 66

1-29IM MX100-01EExplanation of Functions32145Index1.10 Functions of the 8-CH, Medium-Speed PWM Output ModuleThis module is equipped with eight output

Página 67 - Connection Procedure

1-30IM MX100-01E1.11 Operation of the 8-CH Medium-Speed Analog Output Module and the 8-CH Medium Speed PWM Output ModuleThis section provides an expl

Página 68 - Initializing the Settings

iiiIM MX100-01EWARNINGUse the Correct Power Supply Ensure that the source voltage matches the voltage of the power supply before turning ON the power

Página 69 - (instantaneous value)

1-31IM MX100-01EExplanation of Functions32145IndexOutput Operation per Setting Changes (Common)Setting Changes (Contents) Output OperationChanges t

Página 70

1-32IM MX100-01ESteady Output OperationOutput Operation While Communication Is ConnectedCommunication Connection Status Output OperationComm. connect

Página 71 - First-Order Lag Filter

1-33IM MX100-01EExplanation of Functions32145IndexOutput When Abnormal, or after Recovery from AbnormalityOutput Operation during Abnormality (by Modu

Página 72 - 2.8 Handling of the CF Card

1-34IM MX100-01EOutput Operation during Abnormality (by Output Setting)Output Setting Abnormal Status Output OperationTrans. output Referenced inpu

Página 73 - Corrective Actions

1-35IM MX100-01EExplanation of Functions32145IndexOutput Operation after Recovery from Abnormality (by Output Setting)Output Setting Abnormality Reco

Página 74 - Communication Errors

1-36IM MX100-01E1.12 Functions of the 10-CH, Medium-Speed Digital Output ModuleThis module outputs ten contact signals based on the alarm output sett

Página 75 - 3.2 Troubleshooting

2-1IM MX100-01EInstallation and Wiring32145IndexChapter 2 Installation and Wiring2.1 Handling PrecautionsThis section describes the precautions to b

Página 76

2-2IM MX100-01E2.2 InstallationInstallation LocationInstall the instrument indoors in the following locations.• Where the temperature is between 0 t

Página 77

2-3IM MX100-01EInstallation and Wiring32145IndexUse on a Desktop or a FloorEach module has feet that can be attached to the base plate allowing them t

Página 78 - 3.3 Calibration

2-4IM MX100-01E2.3 Attaching the Modules WARNINGTo prevent electric shock and instrument breakdown, do not connect the power supply to the main modu

Página 79

ivIM MX100-01EConventions Used in This ManualUnitk Denotes 1000.K Denotes 1024. Example: 5 KB (file size)Safety MarkingsThe following markings are u

Página 80 - 120V MAX CH TO CH

2-5IM MX100-01EInstallation and Wiring32145Index CAUTIONWhen attaching the main module (conventional model) to the base plate, first remove the two s

Página 81

2-6IM MX100-01E2.4 Connecting the Signal WiresTerminal Arrangement Markings on the Terminal CoverA character indicating the terminal function and a t

Página 82

2-7IM MX100-01EInstallation and Wiring32145IndexAttaching and Removing the Terminal BlockThe terminals of the input/output modules in the figure below

Página 83 - Wiring Diagram

2-8IM MX100-01EAttaching Plates with Screw Terminals and Plates with Clamped Terminals for CurrentYou can attach a plate with screw terminals (sold se

Página 84 - 3.4 Maintenance of Parts

2-9IM MX100-01EInstallation and Wiring32145IndexGeneral Precautions When Wiring the Input/Output Signal Wires WARNING• To prevent the possibility of

Página 85 - Communication Commands

2-10IM MX100-01EModule Type Max. Input VoltageMax. Common Mode VoltageBetween ChannelsInput to GroundDCV/TC/DI input module± 10 VDC: Voltage range of

Página 86

2-11IM MX100-01EInstallation and Wiring32145IndexWiring Procedure1. Turn OFF the power to the MX100.2. Loosen the terminal cover attachment screw an

Página 87

2-12IM MX100-01EWiring the Four-Wire RTD Resistance Input Modules ABICResistance, RTDResistance per lead wire of 10 Ω or lessABICVoltageNothing conn

Página 88

2-13IM MX100-01EInstallation and Wiring32145Index• One-Gauge Three-Wire Method ONOFFNo.1No.2No.3No.4No.5A(+V)B(L)C(-V)D( H) No.1ONNo.2ONNo.3OFFNo.

Página 89 - 4.1 Common Specifications

2-14IM MX100-01E• Opposing Two-Gauge Method ONOFFNo.1No.2No.3No.4No.5A(+V)B( L) C(-V)D( H) No.1ONNo.2OFFNo.3ONNo.4OFFNo.5ONRg1Rg2RRRg2Rg1EeRg1Rg2Set

Página 90 - Measurement

vIM MX100-01E32145IndexContentsSafety Precautions ...

Página 91 - Remote RJC

2-15IM MX100-01EInstallation and Wiring32145Index• Four-Gauge Method ONOFFNo.1No.2No.3No.4No.5A(+V)B( L) C(-V)D( H) -B12, -B35No.1OFFNo.2OFFNo.3OF

Página 92

2-16IM MX100-01EWiring the Digital Input Modules (-D05, -D24)Note• With the digital input module, the (–) terminal and unassigned terminals on all ch

Página 93

2-17IM MX100-01EInstallation and Wiring32145IndexWiring the Analog Output Modules External power supply VoltageV+Load+24 Vpower supply(when using cu

Página 94 - General Specifications

2-18IM MX100-01EWiring the Digital Output Modules CNO250 VDC/0.1 A, 250 VAC/2 A, or30 VDC/2 A (resistance load)Main Output SpecificationsContact mod

Página 95

2-19IM MX100-01EInstallation and Wiring32145Index2.5 Power Connection and ON/OFF WARNING• To prevent the possibility of electric shock when wiring,

Página 96

2-20IM MX100-01EWiring the Power Supply Terminal (Power Supply/Cord Basic Specification Code 1W) WARNING• Set up a switch on the power supply line t

Página 97

2-21IM MX100-01EInstallation and Wiring32145IndexTurning ON/OFF the Power SwitchPressing the “I” side of the power switch turns the instrument ON. Pre

Página 98

2-22IM MX100-01E2.6 Connecting the Ethernet CableConnection ProcedureConnectorsConnect the Ethernet cable to the Ethernet port (10BASE-T/100BASE-TX)

Página 99

2-23IM MX100-01EInstallation and Wiring32145IndexChecking the Communication StatusYou can check the status on the two LEDs at the upper-right and lowe

Página 100 - IM MX100-01E

2-24IM MX100-01E2.7 Measures against Noise on the MX100Technical Information described below on measures against noise is available as reference mate

Página 101

viIM MX100-01EContents1.8 Functions of the 10-CH, High-Speed Digital Input Module (-D05, -D24) ... 1-27Measurement Interval ...

Página 102

2-25IM MX100-01EInstallation and Wiring32145IndexOn the MX100, the integral time is determined by the measurement interval setting as shown in the tab

Página 103 - External Dimensions

2-26IM MX100-01EFirst-Order Lag FilterFor noise sources other than power supply noise, the MX100 is equipped with a first-order lag filter having outp

Página 104

2-27IM MX100-01EInstallation and Wiring32145Index2.8 Handling of the CF CardHandling Precautions of the CF CardNote the following points when using t

Página 105

3-1IM MX100-01ETroubleshooting and Maintenance32145IndexChapter 3 Troubleshooting and Maintenance3.1 Error Display on the 7-Segment LED and Correcti

Página 106

3-2IM MX100-01EModule ErrorsThe left and right digits of the 7-segment LED display “E” and an error code, respectively. The LED illuminates. In the ca

Página 107

3-3IM MX100-01ETroubleshooting and Maintenance32145Index3.2 TroubleshootingIf servicing is necessary, or if the instrument is not operating correctly

Página 108 - Specifications

3-4IM MX100-01EThe MX100 can be detected using the Search button, but connection fails.Probable Cause Corrective Action Ref. PageThe IP address is set

Página 109

3-5IM MX100-01ETroubleshooting and Maintenance32145IndexThe measured value is not correct.Probable Cause Corrective Action Ref. PageThe input wiring i

Página 110

3-6IM MX100-01E3.3 CalibrationIt is recommended that the MX100 be calibrated once a year to assure its measurement accuracy. A calibration instrument

Página 111

3-7IM MX100-01ETroubleshooting and Maintenance32145Index• When calibrating the RTD range of the 4-CH, High-Speed Universal Input module 100Vpk MAX

Página 112 - (MX112) Specifications

viiIM MX100-01E32145IndexContents2.5 Power Connection and ON/OFF ...

Página 113 - 57 150.8

3-8IM MX100-01E• When calibrating the DC voltage range of the 30-CH, Medium-Speed DCV/TC/DI Input module Input terminal of CH2Short the input ter

Página 114 - (MX115) Specifications

3-9IM MX100-01ETroubleshooting and Maintenance32145Index• When calibrating the RTD or resistance range of the 6-CH, Medium-Speed Four-Wire RTD Resist

Página 115

3-10IM MX100-01E For -NDI Connect CH2 onlySwitch the bridge headR2R4R1R3ON 1 2 3 4 512345678OFFSWBridge head(701955 or 701956)SW1OFFSW2OFFSW3O

Página 116 - Handling Abnormal Data

3-11IM MX100-01ETroubleshooting and Maintenance32145IndexCalibration of Temperature Measurements using ThermocouplesRequired Instruments• DC voltage/

Página 117 - (MX120-PWM-M08) Specificaions

3-12IM MX100-01E3.4 Maintenance of PartsThe MX100 does not have parts that need periodic replacement. However, the main module (MX100) uses expendabl

Página 118

3-13IM MX100-01ETroubleshooting and Maintenance32145Index3.5 MX100 Maintenance and Test Using Communication CommandsThe MX100 maintenance/test functi

Página 119 - Specification

3-14IM MX100-01Econ Output connection informationFunction Outputs a list of connection information of the devices connected to the MX100.Setting con

Página 120 - Transmission Output

3-15IM MX100-01ETroubleshooting and Maintenance32145Indexhelp Output command helpFunction Outputs command help. If a command name is specified as a

Página 121 - (MX125) Specifications

3-16IM MX100-01EDescription Outputs settings and statistical information specific to the MX100 in a form enclosed by EA and EN. APP: applalive: AP

Página 122 - 5.1 Overview of Functions

4-1IM MX100-01ESpecification32145IndexChapter 4 Specification4.1 Common SpecificationsStandard operating conditionsOperating temperature range: 0 t

Página 123 - 5.2 Explanation of Functions

viiiIM MX100-01EChapter 5 Dual Save Functions (/DS Option)5.1 Overview of Functions ...

Página 124 - Saving data to the CF Card

4-2IM MX100-01E4.2 Main Module (MX100-E) SpecificationsMeasurementStyle number: S3Meas. range and accuracy: See the measurement range and accuracy

Página 125 - • Normal Operation

4-3IM MX100-01ESpecification32145IndexLinear scaling accuracy: Linear scaling accuracy (digits) = Measurement accuracy (digits) × expansion rate +

Página 126 - • Application Example 2

4-4IM MX100-01ESave mode: The following two modes are available for the save operation when communication is disconnected. Rotary: In this mode, th

Página 127 - CAUTION

4-5IM MX100-01ESpecification32145IndexFile name: MDDIXXXX.MXD M: Month when saving started (local time), 1-9, X (October), Y (November), Z (Decem

Página 128 - • When a Power Cut

4-6IM MX100-01EDisplayLED: Ethernet communication status (LINK, ACT)Two-digit 7-segment LED: Status of the MX100 (unit number, operation status, and

Página 129

4-7IM MX100-01ESpecification32145Index4.3 Base Plate (MX150) SpecificationsNumber of main modules that can be attached: 1 (always attached)Number of

Página 130 - 5.3 CF Card Data File

4-8IM MX100-01E4.4 4-CH, High-Speed Universal Input Module (MX110-UNV-H04) SpecificationsStyle number: S1Type of measurement: DC voltage, thermocoup

Página 131

4-9IM MX100-01ESpecification32145IndexThe inputs indicated below can be used on MXLOGGER or the MX100/DARWIN API (software sold separately).InputMeasu

Página 132

4-10IM MX100-01EReference junction compensation accuracy: When measuring temperature greater than or equal to 0°C and when the temperature of the inp

Página 133

4-11IM MX100-01ESpecification32145IndexPower consumption: Approximately 3 WExternal dimensions: Approximately 57 (W) × 131 (H) × 151 (D) mm (includi