<6. Operation>
6-3
IM 01C27H01-01EN
Using the Device Conguration Tool
Refer to subsection 7.3.14 “Zero-point
Adjustment and Span Adjustment”.
(2) When you cannot obtain the Low Range
Value from the actual measured value of
0%;
Adjust the transmitter output to the actual measured
value obtained by a glass gauge, etc.
[Example]
The measuring range of 0 to 2m; the actual
measured value of 0.8m
0.8
2
Actual measured value=
x100=40.0%
Using the transmitter’s zero-adjustment
screw
Turn the zero adjustment screw to match the
output signal to the actual measured value.
Using the Device Conguration Tool
Refer to subsection 7.3.14 ”Zero Point
Adjustment and Span Adjustment”.
6.3 Starting Operation
After completing the Zero Point Adjustment,
follow the procedure below to start operation.
1) Conrm the operating status.
2) After conrming the operating status, perform
the following:
IMPORTANT
Close the terminal box cover and the
amplier cover. Screw each cover in tightly
until it will not turn further.
Tighten the zero-adjustment cover mounting
screw to secure the cover.
6.4 Connecting to the Field
Wireless Network
Preparation work prior to connecting to a
Field Wireless Network
This transmitter does not need to be connected
with a physical wire. Instead of physical wiring, it
is necessary to set the eld wireless devices to
communicate with before installing the transmitter.
This procedure is called a provisioning.
This transmitter supports provisioning via infrared
communication using a provisioning device and
can be securely connected to a network. If the
provisioning information is not set, the transmitter
cannot be connected to the eld wireless network.
Provisioning:
Provisioning is work to set the security and
network information to enable the transmitter
to be connected to the eld wireless network.
This transmitter supports a provisioning method
using infrared communication.
For details on provisioning using a provisioning
device, and procedure for connecting the
transmitter to the Field Wireless Network, refer
to the User’s Manual, FieldMate Versatile Device
Management Wizard (IM 01R01A01-01E),
YFGW710 Field Wireless Integrated Gateway (IM
01W01F01-01EN) and YFGW410 Field Wireless
Management Station (IM 01W02D01-01EN).
F0605.ai
within 30cm
Figure 6.5 Provisioning Example