IM 12D08B02-01E
Parameter setting 5-15
Code Display Function Function detail X Y Z Default values
mA Outputs
30 Not used
31 *OUTP.F mA output functions Linear 0 0 Linear
Table 1
32 *BURN Burn function No burnout 0 0 No Burn.
Burnout downscale 1
Burnout upscale 2
Pulse burnout 3
33, 34 Not used
35 *TABLE Output table for mA
*0% Linearization table for mA in 5% steps.
*5% The measured value is set in the main
*10% display using the >, ^, ENT keys, for
... each of the 5% interval steps.
... Where a value is not known, that value
*95% may be skipped, and a linear interpolation
*100% will take place.
36-39 Not used
mA