IM 12B07C05-01E4th EditionUser’sManualModel PH450G [Style: S2]pH and ORP Converter

Página 2

2 IM 12B07C05-01E2. GENERAL SPECIFICATIONS OF EXAxt PH450GA) Inputs specs :Dualhighimpedanceinput(≥1012Ω) (under referance conditions)B) In

Página 3

3IM 12B07C05-01E2 GENERALSPECIFICATIONSF) Contact input : Remote wash cycle start.G) Temperature compensation Function :Automaticormanual C

Página 4

4 IM 12B07C05-01ECAUTION This instrument is a Class A product, and it is designed for use in the industrial environment. Please use this instrument i

Página 5 - □-A version) ...16

5IM 12B07C05-01E3 INSTALLATIONANDWIRING3-1. Installation and dimensions3-1-1. Installation siteThe EXAxt 450 converter is weatherproof and can be i

Página 6

6 IM 12B07C05-01Emin.185 (7.25")min.195(7.75")138(5.43")138(5.43")M6M6M5138(5.43")+1 0138(5.43")+1 0Figure 3-2. Opt

Página 7

7IM 12B07C05-01E3 INSTALLATIONANDWIRING3-2. Wiring3-2-1. PreparationRefer to figure 3-4. The relay contact terminals and power supply connections a

Página 8

8 IM 12B07C05-01EHigh voltage sectionContact(S1, S2)outputcablesmAcablesContact(S3, S4)outputcablesSensorCablesInputcontactPowercableSuitable for cabl

Página 9 - PH/ORP TRANSMITTER

9IM 12B07C05-01E3 INSTALLATIONANDWIRING3-3. Wiring the power supply3-3-1. General precautionsMake sure the power supply is switched off. Also, make

Página 10

10 IM 12B07C05-01E3-3-5. Grounding the housingForthesafetyoftheuserandtoprotecttheinstrument against interference, the housing must always b

Página 11 - GENERALSPECIFICATIONS

11IM 12B07C05-01E3 INSTALLATIONANDWIRING3-4. Wiring the contact signals3-4-1. General precautionsThe contact output signals consist of volt-age-fre

Página 12

CommissioningNoteThis page may be referred to when reading pages where subsequent submenu screens are shown in the text. Connection to the relevant su

Página 13 - 3. INSTALLATION AND WIRING

12 IM 12B07C05-01ETable 3-1. Impedance measuring jumpersJumper Settings Application & Sensor Connections1613 17 15NormalpHsensors(includingFU2

Página 14

13IM 12B07C05-01E3 INSTALLATIONANDWIRING3-7. Sensor wiringRefer to figure 3-10, which includes drawings that outline sensor wiring.The EXAxt 450 ca

Página 15 - 3 INSTALLATIONANDWIRING

14 IM 12B07C05-01E* ***GreenYellowRedRedBlueBlueBlueRedBlackRedCable markers11 Temperature12 Temperature13 Reference14 Solution ground15 Glass (meas

Página 16 - High voltage section

15IM 12B07C05-01E3 INSTALLATIONANDWIRINGFigure 3-10.c. Sensor wiring for redox measurement11 TEMPERATURE12 TEMPERATURE13 REFERENCE14 SOLUTION

Página 17 - Unit: mm(inch)

16 IM 12B07C05-01ERemovestandard tuleFigure 3.11.b. Content of grommet setFigure 3.11.a. Grommet set use3-7-2. Sensor cable connection with special g

Página 18 - 100-240 VAC/15 VA/ 50/60Hz

17IM 12B07C05-01E3 INSTALLATIONANDWIRING3-7-3. Sensor cable connections using junc-tion box (BA10) and extension cable (WF10)Where a convenient in

Página 19

18 IM 12B07C05-01EExtension cable may be purchased in bulk quantities or in pre-finished lengths. In the case of bulk quantities cut to length, then i

Página 20

19IM 12B07C05-01E4OPERATIONOFEXAxtPH450G4-1. Main display functionsT ag:EXAxt PH45 0 HOLD W AS H 25.0 236.0 mV S4 S3 S2 S1 Go to trend screenGo to

Página 21

20 IM 12B07C05-01Eofthesensor(default7.00pH).TheZEROvalue indicates the condition of the sensor. If the value exceeds +/- 120 mV (or user defi

Página 22

21IM 12B07C05-01E4OPERATIONOFEXAxtPH450G4-3-11. HART ID =apartoftheHARTdeviceID(descriptor)4-3-12. Software revision =therevisionlevel

Página 23 - IM 12B07C05-01E

PREFACEElectrostatic dischargeThe EXAxt converter contains devices that can be damaged by electrostatic discharge. When servicing this equipment, plea

Página 24 - □-A version)

22 IM 12B07C05-01E4-7. Navigation of the menu structure“HOMEKEY”returntomaindisplay“RETURNKEY”exittopreviousdisplayMain displayPrimary setup

Página 25 - / CONVERTER

23IM 12B07C05-01E4OPERATIONOFEXAxtPH450G

Página 26

24 IM 12B07C05-01E5-1. Sensor setupSensortype:Thesensorconnectiontotheter-minals is determining the setting of this parameter. Three selec-tio

Página 27 - 9.10 pH

25IM 12B07C05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.Manual ManualTemp. 25ºC,77ºF -30ºC,-22ºF

Página 28

26 IM 12B07C05-01E5-5. Calibration settingsCalibration settings for a pH converter involve slope (sensitivity), zero (aspot) and ITP (iso thermal poin

Página 29 - OPERATIONOFEXAxtPH450G

27IM 12B07C05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.Limitsandtiming Zerohigh 120mV 0mV  53

Página 30

28 IM 12B07C05-01E5-7. mA output setupThe general procedure is to first define the function (control, output, simulate, off) of the output and second

Página 31

29IM 12B07C05-01E5MENU STRUCTURECOMMISSIONINGmA2 similar structure to mA1Menu Parameter Default Range values min. max.mA1 (control) Expir

Página 32

30 IM 12B07C05-01EExpire timeIftheoutputisover100%forlongerthantheexpiretime,theoutputwillreturnto0%.Damping timeThe response to a st

Página 33

31IM 12B07C05-01E5MENU STRUCTURECOMMISSIONINGS2, S3, S4 Similar structure to S1 Menu Parameter Default Range values min. max.PID-control

Página 34

ii The following safety symbols are used on the product as well as in this manual.DANGERThis symbol indicates that an operator must follow the instruc

Página 35

32 IM 12B07C05-01E5-9. FailA fail contact is energized when a fail situation occurs.Failsituationsareconfiguredinsection5-10.ForSOFTFailsth

Página 36

33IM 12B07C05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.Generic wash settings Interval time 6 hour 0.

Página 37

34 IM 12B07C05-01E5-11. Error configurationErrors 1/3 ~ 3/3Errors are intended to notify the user of any unwanted situations. The user can determine w

Página 38

35IM 12B07C05-01E5MENU STRUCTURECOMMISSIONING

Página 39

36 IM 12B07C05-01E5-13. Advanced setupDefaultsThe functionality of the EXAxt allows to save and load defaults to come to a known instru-ment setting.

Página 40

37IM 12B07C05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values Low HighHart Network address 0 0 15

Página 41 - MENU STRUCTURECOMMISSIONING

38 IM 12B07C05-01E5-14. Display setupMain displayThe main display consists of three lines with Process Values. Each line is user definable with the re

Página 42

39IM 12B07C05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values Low HighY-axis pH low 0 pH -inf +infY-axis pH high

Página 43

40 IM 12B07C05-01E6-1. Calibration check with buffer solutions.The following tips will help to produce a good calibration.1. Before starting a calibra

Página 44

41IM 12B07C05-01E6CALIBRATIONWhen the right buffer tables are entered, automatic calibration is the easiest and most reliable calibration method to u

Página 45

IM 12B07C05-01EiiiIM 12B07C05-01E4th Edition: Mar. 2012(YK) All Rights Reserved, Copyright © 2007, Yokogawa Electric CorporationTABLE OF CONTENTSPREFA

Página 46

42 IM 12B07C05-01ETermination of a wash cycleThe user can decide to terminate the current wash cycle. This is done in the main screen (all other scree

Página 47

43IM 12B07C05-01E7 MAINTENANCE7-1. Periodic maintenance The converter requires very little periodic main-tenance, except to make sure the front windo

Página 48 - 6. CALIBRATION

44 IM 12B07C05-01EWhere a refillable (flowing electrolyte) refer-ence system is employed, make sure that the reservoir is kept topped up. The rate of

Página 49 - CALIBRATION

45IM 12B07C05-01E8TROUBLESHOOTING8-1. General The EXAxt is a microprocessor-based con-verter that performs continuous self-diagnostics to verify that

Página 50 - AGED SENSOR

46 IM 12B07C05-01E9. QUALITY INSPECTIONQualityInspectionStandards PH450GpH and ORP Converter QIS 12B07C05-01E1st Edition: Jul. 2007 1. Scope This

Página 51 - 7. MAINTENANCE

47IM 12B07C05-01E9QUALITYINSPECTION2/3 QIS 12B07C05-01EThis test is done on the “HIF” display of “Factory Mode”. a. Touch the [Setup] icon. b. Touc

Página 52

48 IM 12B07C05-01E3/3 QIS 12B07C05-01ETable 4 Data Display Current Output (mA DC) 4 4 ±0.02 12 12 ±0.02 20 20 ±0.02 After all tests are completed

Página 53 - 8. TROUBLESHOOTING

49IM 12B07C05-01E9QUALITYINSPECTION

Página 54 - Standards

50 IM 12B07C05-01E10. SPARE PARTSSeeCustomerMaintenancePartsList.

Página 55 - QUALITYINSPECTION

51IM 12B07C05-01EAAPPENDICESAPPENDICESAppendix 1, Buffer tablesNIST (IEC 60746-2)/DIN 19266°C 0 5 10 15 20 25 30 35 38 40 45 50 55 60 70 80 90 951,

Página 56 - QIS 12B07C05-01E

IM 12B07C05-01Eiv 4-3-5. Sensor mV. ...204-3-6. Reference im

Página 57

52 IM 12B07C05-01EOnline menu Level 1 menu Level 2 menu Level 3 menu Level 4 menuProcess values Primary value (pH)*Secondaryvalue(Temp.)Tertiaryval

Página 58 - 10. SPARE PARTS

53IM 12B07C05-01EAAPPENDICESOnline menu Level 1 menu Level 2 menu Level 3 menuLevel 4 menuMost appl. ErrorCalib / WashError description / remedypH 1pt

Página 59 - APPENDICES

54 IM 12B07C05-01EOnline menu Level 1 menu Level 2 menu Level 3 menu Level 4 menuCommissioningMeintenance setup Calib. settings Buffer setSero/Slope/I

Página 60

55IM 12B07C05-01EAAPPENDICESOnline menuLevel 1 menu Level 2 menu Level 3 menu Level 4 menuCommissioning Output setup HOLD setup HOLD L/FmA1fixed*mA2

Página 61

56 IM 12B07C05-01EThe matrix can be cleared before entering new values. Next new matrix values can be entered as described above. The EXAxt can interp

Página 62

57IM 12B07C05-01EAAPPENDICESAppendix 4, Control drawing for FM approval

Página 63

58 IM 12B07C05-01E

Página 64

CMPL12B07C05-02ECustomerMaintenanceParts List©Copyright2007,4thEdition:Oct,2011(YK)Subjecttochangewithoutnotice.Model PH450G [Style: S2]pH

Página 65

CMPL12B07C05-02E24thEdition:Oct.2011(YK)2345311078963Pipe/Wall Mounting Hardware (Option code: /U)Awning Hood (Option code: /H5)Panel Mounting H

Página 66

Revision RecordManual Title : Model PH450G[Style:S2]pHandORPConverter Manual Number : IM 12B07C05-01E Edition Date Remark (s)1st Jul. 2007

Página 67 - Parts List

IM 12B07C05-01Ev8. TROUBLESHOOTING ...458-1. General ...

Página 69 - Revision Record

1IM 12B07C05-01E1 INTRODUCTIONANDGENERALDESCRIPTIONThe Yokogawa EXAxt PH450G is a converter designed for industrial process monitoring, meas-ureme