IM 12B6J3-02E-E
7th edition
Instruction
Manual
Model FU20
pH/ORP Combination sensor
Vista de página 0
1 2 3 4 5 6 ... 11 12

Resumo do Conteúdo

Página 1 - Instruction

IM 12B6J3-02E-E 7th editionInstructionManualModel FU20pH/ORP Combination sensor

Página 2 - TABLE OF CONTENTS

IM 12B6J3-02E-E85-1. Calibration for pH measurementTo calibrate a pH sensor, two buffer solutions with known pH values are required. It is recommended

Página 3 - 1. GENERAL

IM 12B6J3-02E-E96. MODEL CODESThe FU20 has a reference system of Saturated Silver/Silver Chloride and the values are given in the following table as f

Página 4

IM 12B6J3-02E-E Printed in The Netherlands, 07-1004 (A) ISubject to change without notice Copyright ©7. SPARE PARTSSpare part DescriptionFU

Página 5 - 3. INSTALLATION OF FU20

TABLE OF CONTENTS1. General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Página 6 - IM 12B6J3-02E-E

IM 12B6J3-02E-E11. GENERAL1-1. IntroductionThis instruction manual provides information for the installation and use of the FU20, four-in-one wide bod

Página 7 - FU20 + ADAPTERFU20 + FD40

IM 12B6J3-02E-E22. GENERAL SPECIFICATIONS FU202-1. Measuring elements : pH glass electrode : Silver chloride reference : Solid platinum electrode

Página 8 - )68.0( 0.22 Ø

IM 12B6J3-02E-E33. INSTALLATION OF FU20For optimum measurement results, the FU20 should be installed in a location that offers an acceptable represen

Página 9 - 4. WIRING

IM 12B6J3-02E-E4PG7PG7PG7PG7PG7Wrench FlatsTeflon TapeProcess FlowDo NOT overtightenFigure 2. Sensor & adapter InstallationTeflon Tape

Página 10

IM 12B6J3-02E-E5FU20O-ringFittingAdapter( /FPS )O-ringPG7PG7PG7/HCNF, SPRAY CLEANERFU20 + ADAPTERFU20 + FD40Figure 3. Installation examples for the FU

Página 11 - 6. MODEL CODES

IM 12B6J3-02E-E63/4” TAPERED THREAD3/4” TAPERED THREAD22.5 (0.89”)19.7 (0.78”)19.7(0.78”))”20.1( 0.62 Ø1” NPT or ISO 7/1-R151.0 (2.3”)150.0 (5.9”)10.0

Página 12 - 7. SPARE PARTS

IM 12B6J3-02E-E74. WIRING(pH) A(Ref.)B(pH shield) C(ORP/LE)D(Temp)E(Temp)FVP-connector4.1. Cable marking and pin allocation. FU20 cableWire color WU