15
IM WX1-01E
9. Message and Corrective Actions
Error
No. Message Corrective Actions
E211 Cannot write to file. Check if the disk capacity is sufficient or if the file systems
is normal.
E212 Cannot read file. Check if the file exists and is supported by the software or
if the file system is normal.
E213 Cannot open file. Check if the file exists and is supported by the software or
if the file system is normal
E401 Communication error. Check if the recorder connected for communication
is powered on and if the cable is properly connected.
Also check the following items according the the
communication type.
• For Ethernet
Check if address settings are correct; the TCP/IP protocol
is installed in Windows; the Ethernet card is properly
installed.
• For RS-232 and RS-422-A
Check if the baud rate settings match; the port (COM1 to
COM9) settings match, the address settings are correct
(RS-422-A); the serial port of the PC is active and the
appropriate cable is being used.
E402 Communication timeout. -
E403 Cannot open a communication port. Same as E401.
E501 Invalid license number. Please reinstall the software. Install the software again.
E1010 Execution of a process failed. Check whether an executable function exists, or whether
its files are damaged. If this error appears frequently,
reinstall the software.
E1011 Execution of a service failed. Check whether an executable function exists, or whether
its files are damaged. If this error appears frequently,
reinstall the software.
Message
No. Mesage
M1201 Model determination was successful.
M1210 Setting changes saved before execution.
Comentários a estes Manuais