2-19
IM WX103-01E
Logger
Explanation
Input Mode and Measurement Range
The measurement channel settings differ depending on the combination of the
MXLOGGER software version and the MX100 hardware style number. The table below
shows the relationship between MXLOGGER measurement channel settings (by
software version) and the MX100 hardware style.
MXLOGGER MX100 Hardware
Style Number
*2
Notes
Software version
*1
Connectable module types
or setting items
S3 S2 S1
R2.05 or later 30ch module
MX110-VTD-L30
Yes No No
GOST range Yes No No Can be set only for MX110-UNV-M10,
MX110-UNV-H04, MX110-V4R-M06, and
MX110-VTD-L30
Y
es: Setting and operation allowed No: Setting possible, but not operation
*1 T
o check the version, see section 2.16, “Displaying the Software Version and Other
Information.”
*2 The style number is printed on the MX100 main module name plate.
On the channels of the universal input module, select from the modes below according
to the input signal. For details of the measurable range, measurement accuracy, and
resolution, see chapter 4, “Specifications” in the MX100 Data Acquisition Unit User’s
Manual (in the manual CD-ROM)
•
VOLT (DC voltage)
Select the measurement range from 20 mV
, 60 mV, 200 mV, 2 V (default), 6 V
, 20 V,
100 V, 60 mV (HQ), 1 V (HQ), and 6 V (HQ). HQ denotes high resolution.
Note
Current input is possible by attaching a shunt resistor to the input terminal and converting to voltage
input. The shunt resistors in the table below are available for purchase separately. For example, a
250 Ω shunt resistor is used to convert the signal in the range of 1 to 5 V for 4 to 20 mA input.
Part Name Model Resistance
Shunt resistor 415920 250 W ± 0.1%
(for clamp terminal) 415921 100 Ω ± 0.1%
415922 10 Ω ± 0.1%
Shunt resistor 438920 250 Ω ± 0.1%
(for screw terminal) 438921 100 Ω ± 0.1%
438922 10 Ω ± 0.1%
• TC (Thermocouple)
Select the thermocouple type (referred to as Range in the setup) from Type-R (default), Type-S,
Type-B, Type-K, Type-E, Type-J, Type-T, Type-N, Type-W, Type-L, Type-U, and KpvsAu7Fe,
PLATINEL, PR40-20, NiNiMo, WRe3-25, W/WRe26, Type-N(AWG14), and Type-XK.
• RTD1 (resistance temperature detector, measurement current: 1 mA) on the 4-CH,
High-Speed Universal Input Module
Select the RTD type (referred to as Range in the setup) from Pt100 (default), JPt100,
HQ Pt100, HQJPt100, Ni100:SAMA, Ni100:DIN, Ni120, RN Pt100, RN JPt100 and
Pt100G. HQ and RN denote high resolution and noise resistance, respectively.
2.4 Setting the Measurement Conditions (Measurement Channel Settings)
Comentários a estes Manuais