Yokogawa Button Operated DX2000 Manual do Utilizador Página 11

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 21
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 10
10
3. Accuracy and Display Resolution
3.1. Measurement and Display Accuracy (Digital Display)
Computing method of the display value accuracy
¾ Input voltage X
¾ Lower limit of display span VL Upper limit of display span VU
¾ Scale lower limit SL Scale upper limit SU
¾ Digital display value Y
1) Converting equation for the input voltage (setting)
Y = 10
(SU-SL)*(X-VL)/(VU-VL)+SL
2) Measurement accuracy of input voltage X (uses the negative side due to the
LOG characteristic)
X
err
=X – (measurement accuracy of the voltage range)
3) Hardware error of the display value
Y’= Y – Y
err
The difference from the true value Y is hardware error Y’ when display
value Y
err
corresponds to input voltage X
err
.
4) Display accuracy of the mantissa
Display accuracy = Y’* 1.1 + 1 digit
Vista de página 10
1 2 ... 6 7 8 9 10 11 12 13 14 15 16 ... 20 21

Comentários a estes Manuais

Sem comentários