Yokogawa Touch Screen GP20 Manual do Utilizador Página 376

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 375
5-22
IM 04L51B01-01EN
Taking measures
against noise?
NO
YES
• Move the input and communication
wires away from noise sources.
• Ground the recorder.
• Use shielded cables for input and
communication wires.
• Insert an isolation transformer into
the power line.
• Insert a power line filter into the
power line.
• If an inductive load is used in the
alarm output line, insert a serge
suppressor in the line.
Servicing is required. Contact your
nearest YOKOGAWA dealer.
Display and other functions do not work properly.
Are the conditions for
displaying the menu item met?
NO
YES
• Confirm that the GX/GP is equipped
with the necessary options.
Servicing is required. Contact your
nearest YOKOGAWA dealer.
Some display menus are not displayed.
5.2 Troubleshooting
Vista de página 375
1 2 ... 371 372 373 374 375 376 377 378 379 380 381 ... 445 446

Comentários a estes Manuais