Resumo do Conteúdo

Página 1 - Fieldbus Communication

User’sManualYTA Series Temperature TransmitterFieldbus CommunicationIM 01C50T02-01EIM 01C50T02-01E8th Edition

Página 2 - CONTENTS

<3. About Fieldbus>3-2IM 01C50T02-01E3.4 Wiring System CongurationThe number of devices that can be connected to a single bus and the cab

Página 3

<4. Getting Started>4-1IM 01C50T02-01E4. Getting StartedFieldbus is fully dependent upon digital communication protocol and differs in ope

Página 4

<4. Getting Started>4-2IM 01C50T02-01E4.2 Host SettingTo activate Fieldbus, the following settings are required for the host.IMPORTANTDo n

Página 5 - 1. Introduction

<4. Getting Started>4-3IM 01C50T02-01E4.4 Integration of DDIf the host supports DD (Device Description), the DD of the YTA needs to be ins

Página 6 - ■ Warranty

<5. Conguration>5-1IM 01C50T02-01E5. CongurationThis chapter contains information on how to adapt the function and performance of the YT

Página 7 - ■ ATEX Documentation

<5. Conguration>5-2IM 01C50T02-01E5.2 Network DenitionBefore connection of devices with Fieldbus, dene the Fieldbus network. Allocate P

Página 8 - 2. Part Names

<5. Conguration>5-3IM 01C50T02-01E5.3 Denition of Combining Function BlocksThe input/output parameters for function blocks are combined.

Página 9 - 3. About Fieldbus

<5. Conguration>5-4IM 01C50T02-01E5.4 Setting of Tags and AddressesThis section describes the steps in the procedure to set PD Tags and

Página 10 - IM 01C50T02-01E

<5. Conguration>5-5IM 01C50T02-01ETable 5.4 VCR Static EntrySub-indexParameter Description1 FasArTypeAndRole Indicates the type and role

Página 11 - 4. Getting Started

<5. Conguration>5-6IM 01C50T02-01E5.6 Block SettingSet the parameter for function block VFD.5.6.1 Link ObjectLink object combines the da

Página 12 - 4.3 Bus Power ON

1IM 01C50T02-01EYTA seriesTemperature Transmitter Fieldbus CommunicationIM 01C50T02-01E 8th EditionCONTENTS1. Introduction ...

Página 13

<5. Conguration>5-7IM 01C50T02-01E5.6.3 View ObjectThis is the object to form groups of parameters in a block. One of advantage brought b

Página 14 - 5. Conguration

<5. Conguration>5-8IM 01C50T02-01ETable 5.12 View Object for Transducer BlockRelative indexParameterVIEW 1 VIEW 2 VIEW 3 VIEW 4 (1st)VIEW

Página 15 - 5.2 Network Denition

<5. Conguration>5-9IM 01C50T02-01ERelative indexParameterVIEW 1 VIEW 2 VIEW 3 VIEW 4 (1st)VIEW 4 (2nd)VIEW 4 (3rd)VIEW 4 (4th)VIEW 4 (5th)

Página 16 - Function Blocks

<5. Conguration>5-10IM 01C50T02-01ERelative indexParameterVIEW 1 VIEW 2 VIEW 3 VIEW 4 (1st)VIEW 4 (2nd)VIEW 4 (3rd)VIEW 4 (4th)VIEW 4 (5th

Página 17 - 5.5 Communication Setting

<5. Conguration>5-11IM 01C50T02-01ETable 5.13 View Object for AI Function BlockRelative indexParameterVIEW1VIEW2VIEW3VIEW41 ST_REV 2 2 2

Página 18

<5. Conguration>5-12IM 01C50T02-01E5.6.4 Parameters of Transducer BlockThe transducer block makes settings for the temperature transmitte

Página 19 - 5.6 Block Setting

<5. Conguration>5-13IM 01C50T02-01E If you want to switch back to select sensor 1 input while the backup action is active after the senso

Página 20 - 5.6.3 View Object

<5. Conguration>5-14IM 01C50T02-01E CAL_POINT_HI_1 (2), CAL_POINT_LO_1 (2) These parameters store the calibrated upper and lower range l

Página 21

<5. Conguration>5-15IM 01C50T02-01E PV_FTIMEStipulates the time constant (in seconds) of the rst-order lag lter inside the AI block.

Página 22

<6. In-process Operation>6-1IM 01C50T02-01E6. In-process Operation6.1 Mode TransitionAll function blocks have modes. All blocks have the

Página 23

2IM 01C50T02-01E5.6.4 Parameters of Transducer Block ...5-125.6.5 Parameters of AI Function Block

Página 24

<6. In-process Operation>6-2IM 01C50T02-01ETable 6.1 Alert ObjectSubindexParameter NameExplanation Analog AlertDiscreteAlertUpdateAlert1 1

Página 25

<6. In-process Operation>6-3IM 01C50T02-01EFollowing gure shows the items shown on a display.F0603.ai(5)(3)(2)(1)(4)Figure 6.3 LCD Displa

Página 26

<7. Errors and Warnings>7-1IM 01C50T02-01E7. Errors and Warnings7.1 Error and Warning IndicationsFaults found as a result of self-diagnos

Página 27

<7. Errors and Warnings>7-2IM 01C50T02-01ECode Displayed on LCDIndication of DEVICE_STATUS_#Cause RemedyAL032 Not Used Sensor2 Although a c

Página 28

<7. Errors and Warnings>7-3IM 01C50T02-01ETable 7.2 WarningsCode Displayed on LCDIndication of DEVICE_STATUS_#Cause RemedyAL100 AI1 in O/S

Página 29 - 6. In-process Operation

<7. Errors and Warnings>7-4IM 01C50T02-01ECode Displayed on LCDIndication of DEVICE_STATUS_#Cause RemedyAL150 PID2 in Bypass active The byp

Página 30 - Indicator

<7. Errors and Warnings>7-5IM 01C50T02-01E7.3 Checking with DEVICE_STATUS_1 to _8 of Resource BlockWhen faults occur, the corresponding bi

Página 31 - Figure 6.3 LCD Display

<7. Errors and Warnings>7-6IM 01C50T02-01ETable 7.5 Contents of DEVICE_STATUS_2Hexadecimal IndicationIndication When Device Description Ha

Página 32 - 7. Errors and Warnings

<7. Errors and Warnings>7-7IM 01C50T02-01ETable 7.7 Contents of DEVICE_STATUS_5Hexadecimal IndicationIndication When Device Description Ha

Página 33

<7. Errors and Warnings>7-8IM 01C50T02-01ETable 7.9 Contents of DEVICE_STATUS_7Hexadecimal IndicationIndication When Device Description Ha

Página 34

3IM 01C50T02-01EAppendix 4. PID Block ...A4-1A4.1 Function Diagram .

Página 35

<7. Errors and Warnings>7-9IM 01C50T02-01ETable 7.11 Example of Warning Masking SettingsCase Device CongurationsWARNING_ ENABLE_1WARNING_

Página 36

<8. Handling Caution>8-1IM 01C50T02-01E8. Handling CautionThis chapter describes important cautions regarding the installation of explosio

Página 37

<8. Handling Caution>8-2IM 01C50T02-01ENote 2. Wiring* All wiring shall comply with Canadian Electrical Code Part I and Local Electrical Co

Página 38

<8. Handling Caution>8-3IM 01C50T02-01E* When combined with FISCO model IIC barrier Ui = 17.5 V, Ii = 360 mA, Pi = 2.52 W, Ci = 1.5 nF, L

Página 39 - 7.4 Precautions on Warnings

<8. Handling Caution>8-4IM 01C50T02-01ESupply unitThe supply unit must be certied by a notify body as FISCO model and following trapezoida

Página 40

  <8.HandlingCaution>8-5IM 01C50T02-01ESensor circuitEEx ia IIC T4Maximum Voltage (Uo) = 7.7 VMaximum Current (Io) = 70 mAMaximum Power (Po

Página 41 - 8. Handling Caution

  <8.HandlingCaution>8-6IM 01C50T02-01E(2) ElectricalConnectionThe type of electrical connection is stamped near the electrical connectio

Página 42 - 8.1.2 ATEX Certication

<8. Handling Caution>8-7IM 01C50T02-01E *1: The production year The third gure from the left of the serial number shows the year of pr

Página 43

<8. Handling Caution>8-8IM 01C50T02-01E IFM018-A12 Installation Diagram(Intrinsically safe, Division 1 Installation)Non Hazardous Locat

Página 44

<8. Handling Caution>8-9IM 01C50T02-01EThe allowed voltage Uo of the associated apparatus used to supply the bus is limited to the range of

Página 45 - F0804.ai

<1. Introduction>1-1IM 01C50T02-01E1. IntroductionThis manual contains a description of the YTA320 Temperature Transmitter Fieldbus Commun

Página 46 - Location of the marking

<8. Handling Caution>8-10IM 01C50T02-01E*4: Associated Nonincendive Field Wiring Apparatus manufacturer’s installation drawing must be fol

Página 47 - C2G218541

<8. Handling Caution>8-11IM 01C50T02-01E Installation Diagram(Installation Diagram for Intrinsically Safe)Non Hazardous LocationHazardous

Página 48 -  FISCO Rules

<8. Handling Caution>8-12IM 01C50T02-01ESupply unitTrapezoidal or rectangular output characteristic only Uo = 14...24 V (I.S. maximum valu

Página 49

<8. Handling Caution>8-13IM 01C50T02-01EElectrical Data:• Supply Input (+ and –) Maximum Input Voltage Ui: 32 V Maximum Internal Capacita

Página 50 - Proof Type

<9. General Specications>9-1IM 01C50T02-01E9. General Specications9.1 Standard SpecicationsFor items other than those described below,

Página 51

  <9.GeneralSpecications>9-2IM 01C50T02-01E9.2 OptionalSpecicationsForitemsotherthanthosedescribedbelow,refertoIM01C50B01-01E

Página 52 - Terminal

<9. General Specications>9-3IM 01C50T02-01ETable 9.1 AccuracySensor TypeReferenceStandardMeasurement RangeMinimum Span(Recommended)Accura

Página 53

<9. General Specications>9-4IM 01C50T02-01ETable 9.2 Ambient Temperature Effect (/10°C Chang)Sensor TypeInput RangeA/D Coefcient °C °FT/

Página 54 - 9. General Specications

<Appendix 1. List of Parameters for Each Block of the YTA>A1-1IM 01C50T02-01EAppendix 1. List of Parameters for Each Block of the YTANote:

Página 55 - 9.2 OptionalSpecications

<Appendix 1. List of Parameters for Each Block of the YTA>A1-2IM 01C50T02-01ERelative IndexIndex Parameter Name Factory DefaultWrite ModeEx

Página 56 - (Recommended)

<1. Introduction>1-2IM 01C50T02-01E For Safe Use of ProductFor the protection and safety of the operator and the instrument or the system

Página 57

<Appendix 1. List of Parameters for Each Block of the YTA>A1-3IM 01C50T02-01ERelative IndexIndex Parameter Name Factory DefaultWrite ModeEx

Página 58 - Block of the YTA

<Appendix 1. List of Parameters for Each Block of the YTA>A1-4IM 01C50T02-01ERelative IndexIndexParameter NameFactory DefaultWrite ModeExpl

Página 59 - Relative

<Appendix 1. List of Parameters for Each Block of the YTA>A1-5IM 01C50T02-01ERelative IndexIndexParameter NameFactory DefaultWrite ModeExpl

Página 60 - A1.2 Al Function Block

<Appendix 1. List of Parameters for Each Block of the YTA>A1-6IM 01C50T02-01ERelative IndexIndexParameter NameDefault (factory setting)Writ

Página 61

<Appendix 1. List of Parameters for Each Block of the YTA>A1-7IM 01C50T02-01EA1.4 Transducer Block Relative IndexIndex Parameter NameDefau

Página 62 - A1.3 Dl Function Block

<Appendix 1. List of Parameters for Each Block of the YTA>A1-8IM 01C50T02-01ERelative IndexIndex Parameter NameDefault (factory setting)Wri

Página 63 - (factory setting)

<Appendix 1. List of Parameters for Each Block of the YTA>A1-9IM 01C50T02-01ERelative IndexIndex Parameter NameDefault (factory setting)Wri

Página 64 - A1.4 Transducer Block

<Appendix 1. List of Parameters for Each Block of the YTA>A1-10IM 01C50T02-01ERelative IndexIndex Parameter NameDefault (factory setting)Wr

Página 65

<Appendix 1. List of Parameters for Each Block of the YTA>A1-11IM 01C50T02-01ERelative IndexIndex Parameter NameDefault (factory setting)Wr

Página 66

<Appendix 1. List of Parameters for Each Block of the YTA>A1-12IM 01C50T02-01ERelative IndexIndex Parameter NameDefault (factory setting)Wr

Página 67

<1. Introduction>1-3IM 01C50T02-01E ATEX DocumentationThis procedure is only applicable to the countries in European Union.GBDKIENLSFPFDSL

Página 68

<Appendix 2. Parameters for Basic Settings, and How to Make and Change the Settings>A2-1IM 01C50T02-01EAppendix 2. Parameters for Basic Settin

Página 69 - A1.5 Unit and Code

<Appendix 2. Parameters for Basic Settings, and How to Make and Change the Settings>A2-2IM 01C50T02-01ETo Do This Corresponding Parameters Outl

Página 70 - Settings, and How to Make

<Appendix 2. Parameters for Basic Settings, and How to Make and Change the Settings>A2-3IM 01C50T02-01E(1) Making the Input Sensor SettingsAcc

Página 71

<Appendix 2. Parameters for Basic Settings, and How to Make and Change the Settings>A2-4IM 01C50T02-01EAccess the parameter CAL_STATE_1 and set

Página 72

<Appendix 2. Parameters for Basic Settings, and How to Make and Change the Settings>A2-5IM 01C50T02-01EAccess the parameter SENSOR_TYPE_1 and s

Página 73

<Appendix 2. Parameters for Basic Settings, and How to Make and Change the Settings>A2-6IM 01C50T02-01E(5) Setting the Low Cut-off LevelSet th

Página 74 - A2.4 Setting Up AI Blocks

<Appendix 3. Function Block Diagram>A3-1IM 01C50T02-01EAppendix 3. Function Block Diagram A3.1 AI Block Function DiagramFA0301.aiAITransd

Página 75 - A2.5 Setting Up DI Blocks

<Appendix 4. PID Block>A4-1IM 01C50T02-01EAppendix 4. PID BlockA PID block performs the PID control computation based on the deviation of

Página 76 - DITransducer OUT_D

<Appendix 4. PID Block>A4-2IM 01C50T02-01EA4.3 Parameters of PID BlockNOTE: In the table below, the Write column shows the modes in which

Página 77 - Appendix 4. PID Block

<Appendix 4. PID Block>A4-3IM 01C50T02-01EIndex Parameter NameDefault (factory setting)Write Valid Range Description32 RCAS_IN 0 Remote se

Página 78 - A4.3 Parameters of PID Block

<2. Part Names>2-1IM 01C50T02-01E2. Part NamesRefer to the individual instruction manuals for detailed descriptions of the parts. This sec

Página 79

<Appendix 4. PID Block>A4-4IM 01C50T02-01EIndex Parameter NameDefault (factory setting)Write Valid Range Description61 HI_ALM — — As abo

Página 80 - A4.5 Control Output

<Appendix 4. PID Block>A4-5IM 01C50T02-01EA4.5.1 Velocity Type Output ActionThe PID block determines the value of the new control output O

Página 81 - A4.9 Block Modes

<Appendix 4. PID Block>A4-6IM 01C50T02-01EBlock Mode DescriptionROut Remote output mode, in which the PID block outputs the value set in RO

Página 82 - A4.11 Setpoint Limiters

<Appendix 4. PID Block>A4-7IM 01C50T02-01ESetpoint Rate LimitsThe setpoint rate limits are used to restrict the magnitude of changes in the

Página 83

<Appendix 4. PID Block>A4-8IM 01C50T02-01EA4.14 Initialization and Manual Fallback (IMan)Initialization and manual fallback denotes a set

Página 84 - A4.16 Auto Fallback

<Appendix 4. PID Block>A4-9IM 01C50T02-01EA4.17 Mode Shedding upon Computer FailureWhen the data status of RCAS_IN or ROUT_IN, which is th

Página 85 - A4.18 Alarms

<Appendix 4. PID Block>A4-10IM 01C50T02-01EA4.19 Example of Block ConnectionsPIDBKCAL_IN OUTINAOBKCAL_OUTCAS_INAIOUTFA0406.aiWhen conguri

Página 86 - Connections

<Appendix 4. PID Block>A4-11IM 01C50T02-01ERelative IndexParameter MnemonicVIEW 1VIEW 2VIEW 3VIEW 434 SHED_OPT 135 RCAS_OUT 536 ROUT_OUT 53

Página 87

<Appendix 5. Link Master Functions>A5-1IM 01C50T02-01EAppendix 5. Link Master FunctionsA5.1 Link Active SchedulerA link active scheduler

Página 88 - A5.2 Link Master

<Appendix 5. Link Master Functions>A5-2IM 01C50T02-01EA5.3 Transfer of LASThere are two procedures for an LM to become the LAS: • If the

Página 89 - A5.3 Transfer of LAS

<3. About Fieldbus>3-1IM 01C50T02-01E3. About Fieldbus3.1 OutlineFieldbus is a bi-directional digital communication protocol for eld dev

Página 90 - A5.4 LM Functions

<Appendix 5. Link Master Functions>A5-3IM 01C50T02-01E(3) In the LAS settings of the YTA, set the values of V(FUN) and V(NUN) so that they

Página 91 - A5.5 LM Parameters

<Appendix 5. Link Master Functions>A5-4IM 01C50T02-01EA5.5 LM ParametersA5.5.1 LM Parameter ListThe tables below show LM parameters of a

Página 92

<Appendix 5. Link Master Functions>A5-5IM 01C50T02-01EIndex (SM)Parameter Name Sub-parameter Name (Sub Index)Default Factory SettingAccess

Página 93

<Appendix 5. Link Master Functions>A5-6IM 01C50T02-01EA5.5.2 Descriptions for LM ParametersThe following describes LM parameters of a YTA

Página 94 - (11) PlmeBasicInfo

<Appendix 5. Link Master Functions>A5-7IM 01C50T02-01ESub-indexElementSize [bytes]Descrip-tion1 SlotTime 2 V(ST)2 PerDlpduPhlOverhead 1 V(P

Página 95

<Appendix 5. Link Master Functions>A5-8IM 01C50T02-01E(12) LinkScheduleActivationVariableWriting the version number of an LAS schedule, whi

Página 96 - A5.6 FAQs

<Appendix 5. Link Master Functions>A5-9IM 01C50T02-01EA5.6 FAQsQ1. When the LAS stops, a YTA does not back it up by becoming the LAS. Wh

Página 97 - Revision Information

iIM 01C50T02-01ERevision Information Title : YTA series Temperature Transmitter Fieldbus Communication Manual No. : IM 01C50T02-01EEdition D