3-31
IM WX1-03E
Detailed Description of Functions
3.3 Error Messages and Corrective Actions
Error
No. Message Corrective Actions
E211 Cannot write to file. Check if the disk capacity is sufficient or if the file systems is
normal.
E212 Cannot read file. Check if the file exists and is supported by the software or if the
file system is normal.
E213 Cannot open file. Check if the file exists and is supported by the software or if the
file system is normal
E401 Communication error. Check if the recorder connected for communication is powered on
and if the cable is properly connected. Also check the following
items according the the communication type.
• For Ethernet
Check if address settings are correct; the
TCP/IP
protocol is
installed in Windows; the Ethernet card is properly installed.
• For RS-232 and RS-422-A
Check if the baud rate settings match; the port (COM1 to COM9)
settings match, the address settings are correct (RS-422-A); the
serial port of the PC is active and the appropriate cable is being
used.
E402 Communication timeout. -
E403 Cannot open a communication port. Same as E401.
E501
Invalid license number.
Please reinstall the software.
Install the software again.
E1010 Execution of aprocess failed. Check whether an executable function exists, or whether its
files are damaged. If this error appears frequently, reinstall the
software.
E1011 Execuition of a service failed. Check whether an executable function exists, or whether its
files are damaged. If this error appears frequently, reinstall the
software.
E1600 The WT1000/WT2000 does not support With WT1000/WT2000 series instruments, model determination
can
Message
No. Message
M1201 Model determination was successful.
M1210 Setting changes saved before execution.
Executable Function Messages
No. Message Corrective Actions
W[631] Data Lack Reduce the number of acquired data points or connected
instruments, or lengthen the scan interval.
E[673] Cannot open communication Same as E401.
E[674] Communication error Same as E401.
E[675] Communication time out Same as E401.
E[850] Command Error An error was received from the WT. Check the status of the WT.
E[851] Cannot setting WT Check whether the communication status and connected
instruments matches those specified in the software. If they do
not, perform model determination again.
E[852] Recive Continued Check the communication status.
E[853] Recieve data error Check the communication status.
I[606] Recovery Communication Connection recovered.