Resistance temperature detector (RTD)
RTD Type Display Measurable Range
Pt100:1mA PT1 –200.0 to 600.0°C
Pt100:2mA PT2 –200.0 to 250.0°C
JPt100:1mAJ JPT1 –200.0 to 550.0°C
JPt100: 2mAJ JPT2 –200.0 to 250.0°C
Pt50: 2mAJ PT50 –200.0 to 550.0°C
Ni100:1mA-S
*1
NI1 –200.0 to 250.0°C
Ni100:1mA-D
*2
NI2 –60.0 to 180.0°C
Ni120:1mA
*3
NI3 –70.0 to 200.0°C
Cu10: GE
*4
CU1 –200.0 to 300.0°C
Cu10: L&N
*4
CU2 –200.0 to 300.0°C
Cu10: WEED
*4
CU3 –200.0 to 300.0°C
Cu10: BAILEY
*4
CU4 –200.0 to 300.0°C
Pt100:1mA-H PT1S –140.00 to 150.00°C
Pt100: 2mA-H PT2S –70.00 to 70.00°C
JPt100:1mA-H JPT1S –140.00 to 150.00°C
JPt100: 2mA-H JPT2S –70.00 to 70.00°C
J263*B J263B –0.0 to 300.0K
*1
RTD (SAMA)
*2 R
TD (DIN)
*3 RTD (McGROW EDISON COMPANY)
*4 RTD (Cuid)
Guaranteed accuracy range Cu10: GE –84.4 to 170.0°C
Cu10: L&N –75.0 to 150.0°C
Cu10: WEED –20.0 to 250.0°C
Cu10: BAILEY –20.0 to 250.0°C
Contact point (DI)
Input Type Display Measurable Range
VOLT LEVL 0 to 1
*1
CONTACT CONT 0 to 1
*2
*1 OFF if below 2.4V, ON if above 2.4V
*2 Contact point ON/OFF
mA input (mA)
Measurement Range Display Measurable Range
20mA 20mA –20.000 to 20.000mA
Pulse input (PULSE)
Measurement Range Measurable Range
RATE Determine the count number during 1s interval.
GATE Determine the most occurring state, ON or OFF, during 1s
interval.
Power monitor (AC)
Input Range Display
250V-0.5A 250V-0.5A
250V-5A 250V-5A
25V-0.5A 25V-0.5A
25V-5A 250V-5A
Strain input (STRAIN)
Measurement Range Display Measurable Range
2k 2k –2000.0 to 2000.0µε
20k 20k –20000 to 20000µε
200k 200k –20000 to 20000µε
1.5 DA100 Functions