Vista de página 102
IM 1E10C1-E
11-11
11. PARAMETER LIST
Name Data Range, Units
Default
Value
R/W Description
Indicator
of SE
Pulse scale 0.0000 to 30000 0 W Sets pulse rate. F02
Pulse low cut 0 to 100% 3 W Sets low cut width of pulse output. F03
Pulse width (00) 50% Duty 50% Duty W Selects pulse width. F04
(01) 0.5 ms
(02) 1 ms
(03) 20 ms
(04) 33 ms
(05) 50 ms
(06) 100 ms
Disp select (00) PV % rnge
PV% rnge
W Selects display items on flowmeter indicator. d01
(01) PV
(02) Totl
(03) Rev. totl
(04) Diff. totl
(05) PV %rnge/Totl
(06) PV/Totl
(07) PV %rnge/PV
(08) PV %rnge/Rev.totl
(09) PV/Rev.totl
(10) PV %rnge/Diff.totl
(11) PV/Diff.totl
Fl user sel (00) Not provided Not W Selects whether PV is displayed with user-defined unit. d02
(01) Provided provided
Fl user span 0 to 30000 100 W Sets value dsiplayed in “PV” at 100% output when “Fl user sel” d03
is provided.
Total/pulse (00) No damp Damp W Selects whether instantaneous flow rate values or damped ones n01
(01) Damp are used for totalization and pulse output.
Output mode (00) On active On active W Selects active level of status output set in “Output function” n02
(01) Off active (except “Alarm out”).
Empty pipe (00) Alarm No alarm W Selects alarm effectiveness when flow tube is not filled with n07
(01) No alarm fluid.
Manufacturer Yokogawa
Yokogawa
R Displays “Yokogawa’.
Dev id Its own R Displays device ID.
ID No.
Write protect Yes No R Displays status of “Write Protect”.
No
Universal rev R Displays version of universal commands.
Fld dev rev R Displays version of communication commands for field device.
Software rev R Displays version of software for field device.
Enable wrt 10min Enter characters W Release write protection when the password set in “New
Password” is entered.
New password Enter characters W Sets new password up to 8 characters.
Software seal (00) Keep Keep R Displays software seal status.
(01) Break