7. OPERATION VIA HART COMMUNICATOR
7.2.4 Entering, Setting and Sending Data
The data which are input with the keys are set in the HART Communicator by pressing
ENTER (F4). Then, by pressing SEND (F2), the data are sent to the ADMAG SE. Note
that the data are not set in the ADMAG SE if SEND (F2) is not pressed. All the data set
with the HART Communicator is held in memory unless power is turned off, so every
data can be sent to the ADMAG SE at one lot.
Operation
Entering data on the “Tag” setting display.
On alphabetic characters, only capital letters can be used for setting Tag No. with HART
Communicator.
1. Device setup 3. Basic setup 5. Tag
ADMAG SE :
Tag
HELP DEL ESC ENTER
1
Call up “Tag” setting display.
On the setting display shown above, enter the data as follows:
Example: Set “FIC-1A”.
5. Tag
DEF
8
FF
I
C
-
1
A
GHI
9
F I
ABC
7
F I C
*
: +
F I C -
STU
1
F I C - 1
ABC
7
F I C - 1 A