5. FUNCTION AND DATA SETTINGS
5.5.3 Resetting for Totalization Display
(1) Presetting for Forward Totalization Display
E04, E05 are used for reset or preset the totalization values of the display.
The totalization value presetting enable can be selected at
parameter No. “E04”.
The totalization value presetting enable is selected at
parameter number “E05”. The initial value is “0”, if it is
no setting, the function is zero setting.
During the A30 :TOTAL (totalization value of flow rate)
is shown, the totalization value display is becoming the
value that is set at parameter number “E05”, by touching
SET key twice.
Totalization preset inhibit
Totalization preset enable
Display Description
ContentCode
00
01
Default: 00 (disabled)
Default: 0
SET SHIFT INC
(Twice)
(2) Zero-resetting for Reverse Toralization Display
The totalization value presetting enable can be selected at
parameter No. “E04”.
During the A31 :REV.TOTAL (totalization value of
reverse flow rate) is shown, the totalization value display
is becoming zero, by touching SET key twice.
Totalization preset inhibit
Totalization preset enable
Display Description
ContentCode
00
01
Default: 00 (disabled)
SET SHIFT INC
(Twice)