IM 12J05D02-01E
Parameter setting 5-17
Code Display Function Function detail X Y Z Default values
Temperature Measuring Functions
10 *T.SENS Temp. Comp. Pt100 RTD 0
Pt1000 RTD (DO30) 1 1
PB36 = 2k2 NTC 2
22k NTC (SM31) (DO70G) 3
11 *T.UNIT Temp units Celsius 0 0
Fahrenheit 1
12 *T.ADJ Temp adjust Adjustment +/- 7.5 °C or +/- 13.5 °F
13 *T.MAN Manual TC Automatic temperature comp 0 0
Manual temp compensation 1 25 °C/ 77 °F
14-19 Not used
Code Display Function Function detail X Y Z Default values
Calibration Settings
20 *∆T.SEC
*∆mg/l
Stabilization
Step
Stabilisation time range 10- 600 s 60 s
DO step change range 0-50 mg/l
or 0-300 %
0.05 mg/l
21 *0.CAL Zero cal. Zero calibration disabled 0 0
Zero Calibration enabled 1
22 *ZERO Zero entry Galvanic:
In case that zero check is activated in code 02
within *I.CELL value set in code 02
In case that zero check is not activated in code
02 ±19.99 µA
00.00 µA
000.0 nA
Polarographic:
In case that zero check is activated in code 02
within *I.CELL value set in code 02
In case that zero check is not activated in code
02 ±199.9 nA
*SLOPE Slope entry Minimum setting
(*I.CELL set in code 01)
× (*SL.MIN set in code 02)
Maximum setting
(*I.CELL set in code 01)
× (*SL.MAX set in code 02)
3.75 µA (50 nA)
23-29 Not used