Yokogawa DO402 Dissolved Oxygen Analyzer Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Sensores Yokogawa DO402 Dissolved Oxygen Analyzer. Yokogawa DO402 Dissolved Oxygen Analyzer User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 98
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 0
Users
Manual
Model DO402G [Style: S3]
Dissolved Oxygen Converter
IM 12J05D02-01E
IM 12J05D02-01E
8th Edition
Vista de página 0
1 2 3 4 5 6 ... 97 98

Resumo do Conteúdo

Página 1 - Dissolved Oxygen Converter

User’sManualModel DO402G [Style: S3]Dissolved Oxygen Converter IM 12J05D02-01EIM 12J05D02-01E8th Edition

Página 2

IM 12J05D02-01E1-2 Introduction1-2. ApplicationThe EXA converter is intended to be used for continuous on-line measurement in industrial installations

Página 3

IM 12J05D02-01EIntroduction 1-31-3. General informationFlexibility, reliability and low maintenance are among the benefits provided by the EXA DO402G

Página 4

IM 12J05D02-01E1-4 IntroductionDisplay functions and rangesThe display continuously gives you all necessary information at a glance. The process value

Página 5 - 8th Edition: Feb. 2014(YK)

IM 12J05D02-01EIntroduction 1-5Logbook Software record of important events and diagnostic data. Available through RS485, for use with the Yokogawa PC4

Página 6

IM 12J05D02-01E1-6 Introduction1-5. System Configuration(For DO30G)Output Signal:4 to 20 m A DC or0 to 20 m A DCDO30GDOX8ADOX8 WDO70GDO402GWTB10

Página 7

IM 12J05D02-01EIntroduction 1-7e. Ingold model 32 sensors with 12, 19 and 25 mm shaft These sensors are available in many different configuration

Página 9 - Figure 1-1. Nameplate example

IM 12J05D02-01ESpecification 2-12-1. GeneralA. Input specifications : The DO402G Dissolved Oxygen converter measures the cur-rent, that is ge

Página 10

IM 12J05D02-01E2-2 SpecificationJ. Contact outputs- General : Four (4) SPDT relay contacts with LED indicators. For S1, S2, and S3, the LED is on

Página 11

IM 12J05D02-01ESpecification 2-3B. Performance : Temperature (Pt1000, PB36, 22kNTC)- Linearity : ± 0.3 ºC- Repeatability : ± 0.1 ºC- Accuracy

Página 14 - Calibration Set

IM 12J05D02-01EInstallation and wiring 3-13. INSTALLATION AND WIRING3-1. Installation and dimensions3-1-1. Installation siteThis instrument is a Class

Página 15

IM 12J05D02-01E3-2 Installation and wiringM6, 4 screws100178137+2 0137+2 012 max.(panel thickness)23Panel cutout dimensions(7.01)(0.91)(0.47)(5.43)(

Página 16

IM 12J05D02-01EInstallation and wiring 3-33-2. Wiring3-2-1. Wiring of DO30GDO402G61 62656663 21222313 IE11 T112 T214 15 RE 16 mA 1mA 2Power supplyCo

Página 17

IM 12J05D02-01E3-4 Installation and wiring• Never apply power to the DO402G converter and other instruments connected to the DO402G conv

Página 18 - 한국 전자파적합성 기준

IM 12J05D02-01EInstallation and wiring 3-5Figure 3-6. System configuration3-3. Wiring the power supply3-3-1. General precautionsMake sure the power su

Página 19

IM 12J05D02-01E3-6 Installation and wiring3-3-2. Access to terminal and cable entryTerminals 1 and 2 on the bottom terminal strip are used for the pow

Página 20

IM 12J05D02-01EInstallation and wiring 3-73-3-3. AC powerConnect terminal 1 to the phase line of the AC power and terminal 2 to the zero line. The siz

Página 21 - Unit: mm (inch)

IM 12J05D02-01E3-8 Installation and wiring3-4. Wiring the contact signals3-4-1. General precautionsThe contact output signals consist of voltage-free

Página 22 - 3-2 Installation and wiring

IM 12J05D02-01EInstallation and wiring 3-93-5. Wiring the analog output signals3-5-1. General precautionsThe analog output signals of the EXA transmit

Página 23 - 3-2. Wiring

IM 12J05D02-01EPREFACEElectric dischargeThe EXA analyzer contains devices that can be damaged by electrostatic discharge. When servicing this equipmen

Página 24

IM 12J05D02-01E3-10 Installation and wiring3-7. Wiring other galvanic sensorsConsult the users manual for the color identification of the sensor cable

Página 25

IM 12J05D02-01EInstallation and wiring 3-11DO70G Wiring Diagram ExampleDO402GDOX10Detail-1Detail-2DO70GDetail-1(DO402G)Detail-2(DOX10)Connections of t

Página 27 - (M4 screw)

IM 12J05D02-01EOperation 4-14. OPERATION; DISPLAY FUNCTIONS AND SETTING4-1. Operator interfaceThis section provides an overview of the operation of th

Página 28

IM 12J05D02-01E4-2 OperationFigure 4-1. DO402 operator interface4-2. Explanation of operating keysMODE key This key toggles between the Measuring an

Página 29 - CONT mA OUTPUT

IM 12J05D02-01EOperation 4-34-3. Setting passcodesIn Service Code 52, EXA users can set passcode protection for each one of the three operating levels

Página 30

IM 12J05D02-01E4-4 Operation4-4. Display functions (default)YESYESNOMODENOYESYESNONOYESNOYES NOYES NODISPLAYHOLDAIR.CALH2O.CALNONONONONONONONOYESYESNO

Página 31

IM 12J05D02-01EParameter setting 5-15. PARAMETER SETTING5-1. Maintenance modeStandard operation of the EXA instrument involves use of the maintenance

Página 32

IM 12J05D02-01E5-2 Parameter setting5-1-1. Manual activation of HoldAIR.CALYESNOMODENONONONOYESNOYESHOLDNote: The HOLD feature must first be activated

Página 33

IM 12J05D02-01EParameter setting 5-35-1-2. Manual Wash start/stopNOYESAIR.CALYESNOMODENONONONONONONOYESNOYESYESMODEYESNONote: Wash must first be switc

Página 34 - FRONT EXA DO402

IM 12J05D02-01E This symbol indicates that the operator must refer to the instructions in this manual in order to prevent the instrument (hardware) o

Página 35

IM 12J05D02-01E5-4 Parameter setting5-1-3. Setpoint adjustmentNOYESYESNOMODENOYESYESNOYESNOYESNOYESYESFor adjustments,follow proceduresas in section 5

Página 36 - 4-4 Operation

IM 12J05D02-01EParameter setting 5-55-2. Commissioning modeIn order to obtain peak performance from the EXA converter, you must set it up for each cus

Página 37

IM 12J05D02-01E5-6 Parameter settingENTENTENTENT ENTENTENTENTrepeatedkeystrokesYESNONONONONONONOYESNOYESNOYESYESNOYESNOYESNOYESNOYESNOSETPOINTSRANGESE

Página 38 - 5-2 Parameter setting

IM 12J05D02-01EParameter setting 5-7ENTAdjust setpoint valueusing > ENT keysas shown for setpoint 1.>YESNOYESNOYESNONONONOProcess Alarms onS

Página 39 - WASH ACTIVE press YES

IM 12J05D02-01E5-8 Parameter setting5-2-2. RangeYESNOYESNONONONONONOYESYESNOYESENTYESNOSee facingpageENTENTENTENTENTENTENTENTYESNOYESNOYESNOYESNOSETPO

Página 40 - 5-1-3. Setpoint adjustment

IM 12J05D02-01EParameter setting 5-9YESNOYESNONOYESNONOENTENT ENTENT ENTORYESENTENTYESENTYESNOYESNOYESNOYESNOYESNOChoose Range to adjust, then set b

Página 41

IM 12J05D02-01E5-10 Parameter setting5-2-3. Hold YESNOYES NOYESNOYESNOYESNOYESNOYESNONONONONONOYESYESNOYESNOYESNOYESNOYESYESNOHOLD activelast measured

Página 42 - 5-2-1. Setpoints

IM 12J05D02-01EParameter setting 5-11YESNOENTHOLDHOLDHOLD HOLDHOLDENTHOLDENTENTENTENTENTSet HOLD "fixed value"for mA2.YESSet HOLD "fixe

Página 43 - Parameter setting 5-7

IM 12J05D02-01E5-12 Parameter setting5-2-4. Wash YESNOYESNOYESNOYESNOYESNOYESNOYESNONONONONONOYES YES(NOTE: The menu item "WASH"will not app

Página 44 - 5-2-2. Range

IM 12J05D02-01EParameter setting 5-135-2-5. ServiceENTYESNOYESNOYESNOYESNOYESNONONONONONOSETPOINTSRANGESET HOLDSERVICE*MEASUREMAN.CALDISPLAYHOLDMODEAI

Página 45 - Parameter setting 5-9

IM 12J05D02-01ETABLE OF CONTENTSPREFACE...

Página 46 - 5-2-3. Hold

IM 12J05D02-01E5-14 Parameter setting5-3. Notes for guidance in the use of service coded settings5-3-1. Parameter specific functionsCode 01 *S.TYPE

Página 47 - HOLD HOLD

IM 12J05D02-01EParameter setting 5-15Code Display Function Function detailX Y Z Default valuesParameter Specific functions01 *S. TYPE Sensor typeGalva

Página 48 - 5-2-4. Wash

IM 12J05D02-01E5-16 Parameter setting5-3-2. Temperature functionsCode 10 *T.SENS Select temperature sensor to suit the measuring probe. Refer to in

Página 49 - 5-2-5. Service

IM 12J05D02-01EParameter setting 5-17Code Display Function Function detail X Y Z Default valuesTemperature Measuring Functions10 *T.SENS Temp. Comp. P

Página 50

IM 12J05D02-01E5-18 Parameter setting5-3-4. mA output settingsCode 30 *mA Select 0-20 mA or 4-20 mA for each of the mA outputs.Code 31 *OUTP.F Se

Página 51 - 0 to 99.9%*SL.MIN

IM 12J05D02-01EParameter setting 5-19Code Display Function Function detail X Y Z Default valuesmA Output Functions30 *mA mA output Output 1

Página 52

IM 12J05D02-01E5-20 Parameter setting5-3-5. Contact outputsCode 40, *S1 & *S2 Process relays can be set for a variety of alarm and control func

Página 53 - Parameter setting 5-17

IM 12J05D02-01EParameter setting 5-21Code Display Function Function detail X Y Z Default valuesContact Settings40 *S 1 Contact S1 Contact 1

Página 54

IM 12J05D02-01E5-22 Parameter settingCode 44 *D.TIME The delay time (or dead time) sets the minimum relay switching time. This function can be adj

Página 55 - 51 to 0.1.0

IM 12J05D02-01EParameter setting 5-23Code Display Function Function detail X Y Z Default valuesContacts Settings44 *D.TIME Dead Time Delay a

Página 56

IM 12J05D02-01E5-1-2. Manual Wash start/stop ...5-35-1

Página 57 - Parameter setting 5-21

IM 12J05D02-01E5-24 Parameter settingControl and Alarm FunctionsControl output (mA) : Pl control on the 2nd mA output. The 2nd mA output can be

Página 58

IM 12J05D02-01EParameter setting 5-25100500tontoff50%50%% duty cycle controlDuty cycleton > 0.1 secDuty cycletoff > 0.1 secDuty cycleProportiona

Página 59 - Parameter setting 5-23

IM 12J05D02-01E5-26 Parameter setting5-3-6. User interfaceCode 50 *RET When Auto return is enabled, the converter reverts to the measuring mode fro

Página 60 -

IM 12J05D02-01EParameter setting 5-27Code Display Function Function detail X Y Z Default valuesUser Interface 50 *RET Auto return No return

Página 61 - Interval Time

IM 12J05D02-01E5-28 Parameter setting5-3-7. Communication setupCode 60 *COMM. The settings should be adjusted to suit the communicating device conn

Página 62

IM 12J05D02-01EParameter setting 5-29Code Display Function Function detail X Y Z Default valuesGeneral70 *LOAD Load defaults Reset configura

Página 64

IM 12J05D02-01ECalibration 6-16. CALIBRATION PROCEDURECalibration of the dissolved oxygen analyzer is performed in the following situations:• When a

Página 65 - Parameter setting 5-29

IM 12J05D02-01E6-2 Calibration6-1-2. Diagnostic functions performed during calibrationThe calibration is a semi-automatic calibration, which means th

Página 66

IM 12J05D02-01ECalibration 6-36-2-2. Procedure for air calibrationNOYESMEASUREMAN.CALDISPLAYHOLDNO MODEYESENTYOKOGAWAMODEAIR.CALCONTACTSS1S2WASH/S3FA

Página 67

IM 12J05D02-01E11. Appendix 2 QUALITY INSPECTION ...11-1Customer Maintenance Parts List (

Página 68

IM 12J05D02-01E6-4 Calibration6-3. Calibration procedure using water calibration method6-3-1. PreparationMove the sensor to a maintenance site and wa

Página 69 - Calibration 6-3

IM 12J05D02-01ECalibration 6-56-3-3. Procedure for Water calibrationYES NOYESNOYES NOYESThe instruments waits forthe reading to stabilize.(The displ

Página 70

IM 12J05D02-01E6-6 Calibration6-4. Calibration method using manual calibration method6-4-1. Preparationa. Cleaning Confirm that the readings of the

Página 71 - Calibration 6-5

IM 12J05D02-01ECalibration 6-76-4-2. Procedure for manual calibrationNOYESENTNOYESNOYESYESENTENTENTYESNONOMEASUREMAN.CALDISPLAYHOLDNO MODEYESENTMODEA

Página 72

IM 12J05D02-01E6-8 CalibrationTable 6-3. Solubility of oxygen (mg/l) of sea water & fresh water (based on sea level barometric pressure of 760

Página 73 - Calibration 6-7

IM 12J05D02-01EMaintenance 7-17. MAINTENANCEIt is important for maintaining the measurement accuracy of the EXA DO series of wire dissolvedoxygen mete

Página 74 - 6-8 Calibration

IM 12J05D02-01E7-2 maintenance7-2. Periodic maintenance for the EXA DO402G converterThe DO402G converter requires very little periodic maintenance. T

Página 75

IM 12J05D02-01EMaintenance 7-3How to replace the fuse1) Before replacing the fuse, turn off power to the instrument at the external breaker.2) Remov

Página 77 - Fuse Holder

IM 12J05D02-01ETroubleshooting 8-18. TROUBLESHOOTINGThis chapter describes the countermeasures for failures, classifying the cases into three categor

Página 79

IM 12J05D02-01E8-2 Troubleshooting8-2. Measures in the case of failure (Error) detectionIf a failure is detected through the self-diagnosis of the DO4

Página 80

IM 12J05D02-01ETroubleshooting 8-3 Error Generation Error content and Causes CountermeasuresNo. Mode E10 All modes EEPROM writing failure Tur

Página 82

IM 12J05D02-01ESpare parts 9-19. SPARE PARTSSee Customer Maintenance Parts List.

Página 84

IM 12J05D02-01EAppendix 10-110. APPENDIX Adjustable limits and defaults for settings in Commissioning Level10-1. Setpoint10-2. RangeNote: The range

Página 85

IM 12J05D02-01E10-2 Appendix10-5. User setting table FUNCTION SETTING DEFAULTS USER SETTINGSParameter specific functions01 *S.TYPE 0 Galvanic (7

Página 86 - 10-5. User setting table

IM 12J05D02-01EAppendix 10-3 FUNCTION SETTING DEFAULTS USER SETTINGSUser Interface50 *RET 1 on 51 *MODE 0.0.0 all off 52 *PASS 0.0.0

Página 87 - Appendix 10-3

IM 12J05D02-01E10-4 AppendixError codesCode Error description Possible cause Suggested remedy E1 Response time failure. Calibration failed to Te

Página 88 - Error codes

IM 12J05D02-01EAppendix 2 11-111. APPENDIX 2 QUALITY INSPECTIONQualityInspectionStandards DO402GDissolved Oxygen Converter QIS 12J05D02-01E1st Edi

Página 89 - Standards

IM 12J05D02-01EIntroduction 1-11. INTRODUCTION AND GENERAL DESCRIPTIONThe Yokogawa EXA is a 4-wire coverter designed for industrial process monitoring

Página 90 - 11-2 Appendix 2

IM 12J05D02-01E11-2 Appendix 22/4 QIS 12J05D02-01Ef. Press the [] key seven times. The data display will show “87” with the second digit of 7 flashi

Página 91 - Appendix 2 11-3

IM 12J05D02-01EAppendix 2 11-33/4 QIS 12J05D02-01ETable 3 Temperature (°C) Resistance () Data Display (°C) 5 1019.5 5 ±0.3 25 1097.3 25 ±0.

Página 92 - 11-4 Appendix 2

IM 12J05D02-01E11-4 Appendix 24/4 QIS 12J05D02-01E3.7 Atmospheric Pressure Indication Check Following Section 3.6, press the [ENT] key until the mess

Página 93 - TEST CERTIFICATE

IM 12J05D02-01EAppendix 2 11-5製品名称溶存酸素変換器 タグNo.PRODUCT NAMEDISSOLVED OXYGEN CONVERTERTAG NO.形名MODELDO402G手配No. 計器番号ORDER NO. SERIAL NO.検 査 項 目   INSP

Página 95 - Parts List

CustomerMaintenanceParts ListAll Rights Reserved, Copyright © 2007, Yokogawa Electric Corporation.Subject to change without notice. Model DO402G [Styl

Página 96 - CMPL 12J05D02-03E

CMPL 12J05D02-03E2nd Edition : Sep. 2012 (YK)23453178963Item Part No. Qty Description 1 K9171SS 1 Mounting Set (/U) 2 Y9608KU 4 S

Página 97 - Revision Record

IM 12J05D02-01ERevision RecordManual Title : Model DO402G Dissolved Oxygen Converter [Style: S3] Manual Number : IM 12J05D02-01EEdition Date Rema