<2 OUTILINE>
2-3
IM 01R01A11-01EN
2.2 Verication Item
This Verication Tool has 2 modes, which can be run separately.
1. Standard Verication (Standard VF)
• This mode veries the condition of the AXF without difculty by checking its internal param-
eters. No external instruments are required.
• Some items are run only when FieldMate is in “online (*1)” mode.
2. Enhanced Verication (Enhanced VF) [Optional]
• This mode provides a more reliable verication by performing a detailed overview equivalent
to SERVICE.
• This mode is optional. If only this mode is completed, “Overall Status” is left blank.
• This mode veries AXF by using external instruments, such as Magnetic Flowmeter Calibra-
tor AM012, Handy Calibrator CA150, and so on.
• Even when FieldMate is in “ofine (*2)” mode, this mode can be run; in this case, some set-
tings must be made manually.
There are only 2 possible combinations:
1. Standard VF
2. Standard VF + Enhanced VF
Table 2.2.1 Verication Items
Mode Verication Item Description
Standard VF Circuit (*3)
Magnetic Circuit Checks the Magnetic Circuit.
Excitation Circuit Checks the Excitation Circuit.
Calculation Circuit Checks the Calculation Circuit.
Device Status (*3)
Alarm Check Checks occurring Alarms.
Alarm History Checks historical Alarms.
Physical Appearance
Flowtube Checks the Flowtube appearance.
Converter Checks the Converter appearance.
Enhanced VF Current Output (*4) Checks the Current Output.
Pulse Output (*4) Checks the Pulse Output.
Converter (*4) Checks the Converter Accuracy.
Insulation Resistance
Coil
Checks the Excitation Coil.
(Remote Flowtube Only)
Signal (Electrode)
Checks the Insulation Resistance.
(Remote Flowtube Only)
*1: Online: The status in which the software communicates with a device.
*2: Ofine: The status in which the software does not communicate with a device.
The communication status between FieldMate and AXF can be conrmed on the FieldMate window.
*3: These items are checked only when Verication Tool is in “online” mode.
*4: When Verication Tool is in “ofine” mode, manual setting is necessary.