<6 OPERATION>
6-15
IM 01R01A11-01EN
(A) No Flow (The uid velocity is completely zero)
1) Fill the owtube with uid completely. Be sure that the uid velocity is completely zero.
2
3
1
F060203.ai
Figure 6.2.3 “Flow/ No Flow” Screen
Table 6.2.3 “Flow/ No Flow” Screen
Item
No.
Item Name Description
1 No Flow Selects the “No Flow” test condition.
2 Start Verication (*1)
Runs “Circuitcheck and “Device Status” check.
These checks are performed automatically. The Progress Bar is displayed
at the bottom left.
3 Cancel Stops runningCircuit” check and “Device Status” check.
*1: When the AXF protect function is set to “Protect”, Figure 3.2.2 is displayed. Follow 3.2.2.
2) Select “No Flow (1)” and click “Start Verication (2)”.
3) Before running “Circuit” check and “Device Status” check, Figure 6.2.4 is displayed.