<7 VARIATION OF VERIFICATION DATA>
7-1
IM 01R01A11-01EN
7 VARIATION OF VERIFICATION DATA
In terms of keeping its own data integrity, veri cation data cannot be copied, nor can it be saved
as other name. Veri cation data except for “Locked data” can be saved only by newly saving or
over-writing.
There are 3 types of veri cation data:
1. Installation data
2. Locked data
3. Others
3
2
1
F070101.ai
Figure 7.1.1 Variation of veri cation data
Table 7.1.1 Variation of veri cation data
Item
No.
Item Name Icon Description
1 Installation data
See 7.1.
2 Locked data
See 7.1 and 7.2.
3 Others See 7.3.
7.1 Installation Data
“Installation data” functions as a reference. Each device can have only one “Installation data”.
When the veri cation data is saved by checking the “Save as installation data” box in Figure
6.1.9, the data becomes the Installation data”.
Only when “No Flow + Flow (*1)” is complete, veri cation data can be saved as anInstallation
data”. See 6.1.2.
Save the rst veri cation data after the delivery asInstallation data”. (Recommended)
When the user wants to edit “Installation data” later:
Save the veri cation data by checking only the “Saved as installation databox in Figure
6.1.9. The icon in Data Status is
“.
When it is necessary to save installation data as read-only:
Save the veri cation data by checking the “Saved as installation databox and the Save as
locked data boxin Figure 6.1.9. The icon in Data Status is
”.
*1: “No Flow + Flow”:
“Circuit” check and “Device Status” check for both “No Flow” and “Flow” conditions. Refer to 6.2.1.