Yokogawa PC-Based MX100 Manual do Utilizador Página 59

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 58
2-30
SM MX100-01E
Wiring for bridge voltage accuracy test (with -B12 and -B35)
Dip switch settings: 4-gauge method
ON
1 2 3 4 5
SW1 to SW4: OFF
SW5: ON
Wiring Diagram (Ex.: CH2)
Because the bridge voltage accuracy error will be equal the wiring resistance from the
clamp terminal to the load resistance, the effects of wiring resistance must be kept to
0.5% or less.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
ON
1 2 3 4 5
ON
1 2 3 4 5
ON
1 2 3 4 5
ON
1 2 3 4 5
A/+V
B/ L
C/-V
D/ H
CH1
A/+V
B/ L
C/-V
D/ H
CH2
A/+V
B/ L
C/-V
D/ H
CH3
A/+V
B/ L
C/-V
D/ H
CH4
Load resistance
+
-
Waveform measuring
instrument
Wiring for bridge voltage accuracy test (with -NDI)
Ndis terminal pin number
A
GB
C
F
E
D
Pin name
Signal name
A B C D E F G
P L N H COM PS NS
Wiring Diagram (Ex.: CH2)
CH1
CH2
CH3
CH4
A1002JC
Load resistance
+
-
P
N
PS
NS
Waveform measuring
instrument
2.7 Tests of the 4-CH Strain Input Module