<ESpecicationsCommontoFieldMateBasicandAdvance>
E-62
IM 01R01A01-01E 9th Edition : May 31, 2013-00
Trend Graph Display/operation when Clicking the Start Trend Button,
Change Update Duration Button, and Resume Trend Button are
Performed
Switching between Periodic and One Shot Mode
• SelectingPeriodicfromtheModemenutogglesbetweenthemodes.Themodecanbe
switched from both the Diagnostic Information tab and the Trend Information tab.
• Thefollowingshowsthetrendgraph,dataacquisition,andinternalbufferwhenthescan
mode is switched from periodic to one shot while the Trend Information tab is displayed.
Trend graph: Stopped
Data acquisition: Stopped
Internal buffer storage: Stopped
• Thefollowingshowsthetrendgraph,dataacquisition,andinternalbufferwhenthescan
mode is switched from one shot to periodic.
Trend graph: Displayed
Data acquisition: Resumed
Internal buffer storage: Cleared
Stopping/resuming Trend Operation
• Thefollowingshowsthetrendgraph,dataacquisition,andinternalbufferwhenTrendPause
is selected from the Mode menu to stop trend operation.
Trendgraph:Frozen
Data acquisition: Continues
Internalbufferstorage:Frozen
In other words, the graph is not updated while operation is stopped.
• Thefollowingshowsthetrendgraph,dataacquisition,andinternalbufferwhenTrendPause
is selected from the Mode menu to resume trend operation.
Trend graph: Displayed from time resumed
Data acquisition: Continues
Internal buffer storage: Stored from time resumed
Setting Update Duration
SelectUpdateDurationfromtheSettingmenuwhenthemenuitemisenabled.Themenuitem
is only enabled when periodic mode is set.
Trend Graph Display and Trend Data
(1) ChangingUpdateDuration
Trend graph: Display is refreshed
Internal buffer storage: Cleared
(2) Stopping trend operation
Trend graph: Stopped
Internal buffer storage: Stopped