3-3
IM DR231-11E
Overview and Specifications of RS-422-A/RS-485 Interface
3
3.3 RS-422-A/RS-485 Interface Connection
The following explains how the RS-422-A/RS-485 module is connected to the computer.
Cable Used
There are two types of cables: two-wire cable and four-wire cable. Make sure each type meets the
following conditions.
Cable used : twisted pair shielded cable
2 pairs of 24 AWG minimum (two-wire), 3 pairs 24 AWG mini-
mum (four-wire)
Characteristic impedance : 100 ohm
Capacitance : 50 pF/m
Length of cable : 1.2 km maximum *
* Communication distance of the RS-422-A/RS-485 interface is not the linear distance, but the
total length of the cable (shielded twisted pair cable).
Terminal Arrangement of the RS-422-A/RS-485 Module
RD A Receive data A ( - )
RD B Receive data B ( + )
FG Frame ground
SD A Send data A ( - )
SD B Send data B ( + )
SG Signal ground
Screws used for the terminals:
ISO M4 screws, length 6 mm
Connecting the Cable
Attach crimp-on lugs (for 4 mm screws) with insulation sleeves on the leadwire ends as shown in the
diagram below. Make the exposed portion of the shielded cable to be less than 5 cm.
SD/RD A
SD/RD B
SG
FG
Shield potential
For two-wire system
For four-wire system
SD A
FG
SD B
RD B
RD A
SG
Shield
potential
Short the terminals.
<07. W Title> WARNING
To prevent an electric shock, ensure the main power supply is turned OFF.
Note
As shown on the next page, connect terminal RD to SD(TD) of the computer (converter) and terminal SD to RD of the