6-7
IM DR231-11E
Commands
6
SZ Sets the recording zone.
Mode Operation mode
Setting SZp1, p2, p3<terminator>
p1 Channel number (001 to 560)
Computation channel number (A01 to A60)
p2 Lower limit of zone (0 to 245 mm, or 0 to
145 mm for the DR130)
p3 Upper limit of zone (5 to 250 mm, or 5 to
150 mm for the DR130)
Example Set the recording zone of channel 6 in slot 3 of
subunit 4 to 10 to 50 mm.
SZ436, 10, 50
Comments Set p2 smaller than p3.
This is effective for analog recording.
The optional computation channel number for a
stand-alone model is A01 to A30.
SP Sets partially expanded recording.
Mode Operation mode
Setting SZp1, p2, p3, p4<terminator>
p1 Channel number (001 to 560)
Computation channel number (A01 to A60)
p2 ON/OFF of partially expanded recording
ON Performs partially expanded
recording.
OFF Does not perform partially expanded
recording.
p3 Boundary recording position (1 to 99%)
p4 Boundary value
Example Partially compress and record channel 6 in slot 3 of
subunit 4 at the boundary recording position of 25%
for the boundary value 0 V.
SP436, 25, 0
Comments A waveform can be partially expanded by setting
the set input level (boundary value) as a
percentage of the recording chart (boundary
recording position).
This is effective for analog recording.
If contiguous channels are set to p1, the decimal
point in boundary values when the measurement
ranges for set channels differ take the positions
determined for each measurement range (see
Pages 5-7 and 5-8). If the result exceeds the
measurable ranges, an error occurs. For example,
if the channels of measurement ranges 20 mV, 2
V, and type T thermocouple are set to p1 and the
boundary value is set to 10000, the following
applies:
The boundary value for the channel of the 20
mV measurement range: 10.000 mV
The boundary value for the channel of the 2 V
measurement range: 1.0000 V
The boundary value for the channel of the type
T thermocouple: 1000.0°C
Since the measurement range of the type T
thermocouple is -200.0 to 400.0°C, an error
occurs. For decimal point positions, see Pages
5-7 and 5-8.
The optional computation channel number for a
stand-alone model is A01 to A30.
Set p4 to a value within the recording span, or
within the left/right scale if linear scaling is
applied. It is not possible, however, to set the
recording span nor the left/right scale.
In the case of power monitor module, if the
settings of the wiring method and the input range
are changed, the partially expanded recording
setting is set to OFF.
SG Sets or copies a message.
Message setting
Mode Operation mode
Setting SGp1, p2<terminator>
p1 Message number (01 to 20)
p2 Message character string (Up to 16
characters)
Example Set the message “test” to message No. 5.
SG05, test
Comments For details of the character strings that can be used
in messages, see the characters in the table of
ASCII codes on Page 5-11.
Message copy
Mode Operation mode
Setting SGp1, p2, p3<terminator>
p1 COPY
p2 Message number at a message source (01 to
20)
p3 Message number at a message destination
(01 to 20)
Example Copy message No. 5 to message No. 12.
SGCOPY, 05, 12
ST Sets a tag.
Mode Operation mode
Setting STp1, p2<terminator>
p1 Channel number (001 to 560)
Computation channel number (A01 to A60)
p2 Tag character string (up to 16 characters)
Example Set “TEST” as the tag for channel 6 in slot 3 of
subunit 4.
ST436, TEST
Comments For details of the character strings that can be used
as tags, see the characters in the table of ASCII
codes on Page 5-11.
The optional computation channel number for a
stand-alone model is A01 to A30.
SH Sets a header.
Mode Operation mode
Setting SHp1, p2<terminator>
p1 Setting line number (1 to 5)
p2 Header character string
Example Set the line of number 2 to “TEST RECORDE.”
SH2, TEST RECORDE
Comments Set the character string to be printed in the first
place of a record.
For details of the character strings that can be used
as headers, see the characters in the table of ASCII
codes on Page 5-11.
Parameter p2 can have no more than 60 characters
for theDR130 and no more than 80 characters for
the DR230/240.
SJ Sets a title.
Mode Operation mode
Setting SJp1<terminator>
p1 Title character string (up to 32 characters)
Example Set the title to “TEST RECORDE.”
SJTEST RECORDE
Comments Sets a character string which is recorded in a fixed
interval in analog recording. The interval to be
recorded is set using the XR command (see 6-9).
For details of character strings that can be used as
titles, see the characters in the table of ASCII
codes on Page 5-11.
SF Sets the format for list printing.
Mode Operation mode
Setting SFp1, p2, p3, p4<terminator>
p1 ON/OFF of range list printing (ON, OFF)
p2 ON/OFF of alarm list printing (ON, OFF)
p3 ON/OFF of parameter list related to printing
(ON, OFF)
p4 ON/OFF of other parameters (ON, OFF)
6.4 Setting the Recording