Yokogawa DR240 Manual do Utilizador Página 70

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 121
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 69
6-16
IM DR231-11E
For digital input module
60 mV, 6 V
Example Calibrate the 20-mV range on the module in slot 1 of
subunit 0.
XZ0, 1, CAL/EXEC, 20 mV
Correcting the calibration value manually
(universal module)
Setting XZp1, p2, p3, ..... , p25<terminator>
p1 Unit number (0 to 5)
p2 Slot number (0 to 5)
p3 DISPLAY
p4 Zero calibration value for 20-mV range
p5 Span calibration value for 20-mV range
p6 Zero calibration value for 60-mV range
p7 Span calibration value for 60-mV range
p8 Zero calibration value for 200-mV range
p9 Span calibration value for 200-mV range
p10 Zero calibration value for 2-V range
p11 Span calibration value for 2-V range
p12 Zero calibration value for 6-V range
p13 Span calibration value for 6-V range
p14 Zero calibration value for 20-V range
p15 Span calibration value for 20-V range
p16 Zero calibration value for 50-V range
p17 Span calibration value for 50-V range
p18 Zero calibration value for Pt:1 mA
p19 Span calibration value for Pt:1 mA
p20 Zero calibration value for Pt:2 mA
p21 Span calibration value for Pt:2 mA
p22 Span calibration value for Pt:1 mA-H
p23 Span calibration value for Pt:2 mA-H
p24 Zero calibration value for Cu:2 mA
p25 Span calibration value for Cu:2 mA
Correcting the calibration value manually (DCV/TC/
DI input module)
Setting XZp1, p2, ..... , p17<terminator>
p1 Unit number (0 to 5)
p2 Slot number (0 to 5)
p3 DISPLAY
p4 Zero calibration value for 20-mV range
p5 Span calibration value for 20-mV range
p6 Zero calibration value for 60-mV range
p7 Span calibration value for 60-mV range
p8 Zero calibration value for 200-mV range
p9 Span calibration value for 200-mV range
p10 Zero calibration value for 2-V range
p11 Span calibration value for 2-V range
p12 Zero calibration value for 6-V range
p13 Span calibration value for 6-V range
p14 Zero calibration value for 20-V range
p15 Span calibration value for 20-V range
p16 Zero calibration value for 50-V range
p17 Span calibration value for 50-V range
Correcting the calibration value manually (mA
input module)
Setting XZp1, p2, p3, p4, p5<terminator>
p1 Unit number (0 to 5)
p2 Slot number (0 to 5)
p3 DISPLAY
p4 Zero calibration value for 20-mA range
p5 Span calibration value for 20-mA range
Correcting the calibration value manually (Strain
input module)
Setting XZp1, p2, p3, ....., p9<terminator>
p1 Unit number (0 to 5)
p2 Slot number (0 to 5)
p3 DISPLAY
p4 Zero calibration value for 2k range
p5 Span calibration value for 2k range
p6 Zero calibration value for 20k range
p7 Span calibration value for 20k range
p8 Zero calibration value for 200k range
p9 Span calibration value for 200k range
Correcting the calibration value manually (Digital
input module)
Setting XZp1, p2, ..... , p7<terminator>
p1 Unit number (0 to 5)
p2 Slot number (0 to 5)
p3 DISPLAY
p4 Zero calibration value for 60-mV range
p5 Span calibration value for 60-mV range
p6 Zero calibration value for 6-V range
p7 Span calibration value for 6-V range
Example Modify the span calibration value for the 6V range of
the module in slot 1 of subunit 0 to 32000.
XZ0, 1, DISPLAY, , , , 32000
Storage of A/D calibration data in a module (A/D
calibration END processing)
Setting XZp1, p2, p3, p4<terminator>
p1 Unit number (0 to 5)
p2 Slot number (0 to 5)
p3 END
p4 Storage selection (ABORT, STORE)
Example Store the calibrated values for the module in slot 1 of
subunit 0.
XZ0, 1, END, STORE
Comments You must execute STORE using A/D calibration
end processing each time after making an A/D
calibration or adjustment for one module. If A/D
calibration end processing (STORE) is not
executed, the calibrated data will not take effect.
Since it takes an indefinite time to execute the XZ
command, return an ACK after the processing
ends.
On the controller side, execute the following
processing after receiving an ACK after
transmitting the XZ command. The format of the
ACK status is as follows:
E0Cr+Lf The received command was normally
processed.
E1Cr+Lf There is an error in the received
command.
If A/D calibration data are adjusted, inputting a
numeric value which is out of the setting range
disables the modules to be recognized. For
details, see the separate manual (IM DR231-01E
or IM DR232-01E).
6.5 Other Settings
Vista de página 69
1 2 ... 65 66 67 68 69 70 71 72 73 74 75 ... 120 121

Comentários a estes Manuais

Sem comentários