6-14
IM DR231-11E
KEY_LOCK_ON
Key lock is turned on.
MATH_START
Starts computation.
MATH_CLR_START
Clears computation results then re-start
computation.
MATH_STOP
Stops computation.
MATH_ACK
Clears the computation status indication.
REPORT_START
Starts report.
REPORT_STOP
Stops report.
REP_RECALL_START
Starts printing the report.
REP_PRINT_STOP
Stops printng the report.
KEY_LOCK_OFF
Key lock is turned off.
S/U_LIST_START
Setup list printing starts.
S/U_LIST_STOP
Setup list printing is stopped.
MSG_PRINT
Message printing begins.
ALM_BUF_CLEAR
Alarm buffer is cleared.
MSG_BUF_CLEAR
Message buffer is cleared.
RAM_INIT
Internal memory is initialized.
COMM_INF
Information on communication module
parameter
MODULE_INF
Module information
RE_SYSTEM (DR232/DR242 only)
Reconstruction of a system
ALL_ITEM
All items of parameter p1
p2 Type of screen
OFF No display on the function screen
FUNC1 Screen that appears when the
FUNC key is pressed.
FUNC3 Screen that appears when the
FUNC key is pressed for 3
seconds.
INIT Initialize.(Effective only when p1
is set to ALL_ITEM.)
Example Display setup list printing start on the FUNC1
screen.
XFS/U_LIST_START, FUNC1
Comments • More than one function can be displayed on each
function screen.
•A function can be easily executed by being
displayed on the function screen.
XS Sets which settings are to be
displayed on the setting screen.
Mode Setup mode
Setting XSp1, p2<terminator>
p1 Contents to be displayed on the set screen
SYSTEM: System settings
UNIT Unit settings
MATH Settings for computation of
respective computation
channels
CONST Settings for computation
constants
MEMORY Settings relating to data
saving/reading in the built-
in RAM disk
FLOPPY Settings relating to data
saving/reading on floppy
disks
ZONE Zone recording settings
PARTIAL partially expanded record
settings
CHART2 Second chart speed settings
TAG Tag settings
TIMER Timer settings
LOGIC Logic settings
MESSAGE Message settings
GROUP Group settings
TREND Turning on or off analog
records for each channel
DIGITAL_PR Turning on or off digital
records for each channel
LIST_PR Turning on or off list
printing for each channel
MANUAL_PR Turning on or off manual
printing for each channel
SCALE_PR Turning on or off scale
printing for each channel
ALARM_PR Turning on or off alarm
printing for each channel
HEADER Header settings
TITLE Title settings
MOVE_AVE Moving average settings
INTERPOL Step interpolation settings
MATCH_TIME
Match time settings
LIST_FMT List printing format settings
COPY Copy contents settings
DST Summer/winter time
ALL_ITEM All items of parameter p1
p2 Type of screen
OFF Nothing appears on the set screen.
SET Screen when the set key is pressed
SET3 Screen when the set key is pressed for
3 seconds
INIT Initialize. (Effective only when p1 is
set to ALL_ITEM.)
Example Show the settings of a tag on the screen for which
you have pressed the SET key.
XSTAG, SET
Comments • More than one setting item can be displayed on
the set screen.
• By displaying a setting item on the set screen, the
setting item can be easily set.
• MATH in parameter p1 is valid only for recorders
with the optional computing function or floppy
disk drive function.
• CONST is valid only for recorders with the
optional computing function.
• MEMORY and FLOPPY are valid only for
recorders with the floppy disk drive function.
XB Sets the burnout.
Mode Setup mode
Setting XBp1, p2<terminator>
p1 Channel number (001 to 560)
p2 Selection of burnout
OFF
UP Upscale traveling beyond the scale
DOWN Downscale traveling beyond the scale
Example Set channels 01 to 10 of subunit 0 to upscale
burnout.
XB001-10, UP
Comments • If channels are to be set successively, the setting is
effective only when the channels are in the same
unit.
6.5 Other Settings